NI 5122/5124/5142 Calibration Procedure
|
© National Instruments
|
33
Table 13.
NI 5124 Bandwidth and Flatness Stimuli
and Published Specifications
Iteration
Max Input
Frequency
Input
Impedance
Range
(V)
Input
Frequency
(Hz)
NI 5124
Published
Specifications
Max
Level
(dB)
Min
Level
(dB)
1
150 MHz
50
Ω
0.2
20,100,000
0.6
-0.6
40,100,000
1.5
-1.5
85,000,000
3
†
-3
†
2
150 MHz
50
Ω
0.4
20,100,000
0.5
-0.5
50,100,000
1
-1
100,100,000
1.7
-1.7
150,000,000
3
†
-3
†
3
150 MHz
1 M
Ω
0.2
20,100,000
0.6
-0.6
40,100,000
1.5
-1.5
75,000,000
3
†
-3
†
4
150 MHz
1 M
Ω
0.4
20,100,000
0.5
-0.5
50,100,000
1
-1
100,100,000
1.7
-1.7
145,000,000
3
†
-3
†
5
60 MHz
50
Ω
4
22,100,000
2.0
-1.0
55,000,000
2.0
-1.0
6
20 MHz
50
Ω
0.4
15,100,000
—
-3.5
*
25,100,000
-2.5
*
—
*
Published specifications value listed for validation of noise filter only. The specification is
not included in device specifications documents.
†
Warranted value. Other values are typical. For definitions of warranted specifications and typical
specifications, refer to the
NI PXI/PCI-5124 Specifications
document available online at