NI 5122/5124/5142 Calibration Procedure
|
© National Instruments
|
29
10. Configure the horizontal properties using the niScope Configure Horizontal Timing VI.
11. Repeat steps 11a through 11e for each
Input Frequency
listed for the current iteration in
Table 12 (NI 5122/5142) or Table 13 (NI 5124).
a.
Configure the calibrator to output the
Input Frequency
for the current iteration in
Table 12 (NI 5122/5142) or Table 13 (NI 5124).
b.
Wait 2,500 ms for the impedance matching and frequency of the calibrator to settle.
c.
Initiate a waveform acquisition using the niScope Initiate Acquisition VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_Fetch
Measurement
with the
following parameters:
timeout
:
1.0
vi
: The instrument handle from
niScope_init
channelList
:
"0"
scalarMeasFunction
:
NISCOPE_VAL_VOLTAGE_R
MS
LabVIEW VI
C/C++ Function Call
Call
niScope_Configure
HorizontalTiming
with the
following parameters:
enforceRealtime
:
NISCOPE_VAL_TRUE
numRecords
:
1
vi
: The instrument handle from
niScope_init
minSampleRate
:
100,000,000
(for
NI 5122/5142),
200,000,000
(for NI 5124)
refPosition
:
50.0
minNumPts
:
300,000