NI 5122/5124/5142 Calibration Procedure
|
© National Instruments
|
27
3.
Configure the common vertical properties using the niScope Configure Vertical VI.
4.
Configure the horizontal properties using the niScope Configure Horizontal Timing VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureChan
Characteristics
with the
following parameters:
vi
: The instrument handle from
niScope_init
channelList
:
"0"
inputImpedance
: The
Input
Impedance
value in Table 12
(NI 5122/5142) or Table 13
(NI 5124) for the current
iteration
maxInputFrequency
: The
Maximum Input Frequency
value
in Table 12 (NI 5122/5142) or
Table 13 (NI 5124) for the
current iteration
LabVIEW VI
C/C++ Function Call
Call
niScope_Configure
Vertical
with the following
parameters:
coupling
:
NISCOPE_VAL_DC
probeAttenuation
:
1.0
vi
: The instrument handle from
niScope_init
channelList
:
"0"
range
: The
Range
value in
Table 12 (NI 5122/5142) or
Table 13 (NI 5124) for the
current iteration
offset
:
0.0
enabled
:
NISCOPE_VAL_TRUE