NI 9215 Calibration Procedure

|

© National Instruments

|

5

Figure 2.  

Accuracy Connections for the NI 9215 with BNC

Note

If the calibrator outputs are truly floating, connect the negative output to a 

quiet earth ground as well as COM to give the entire system a ground reference.

3.

Set the calibrator to Operate mode (OPR).

4.

Set the calibrator voltage to a Test Point value indicated in Table 4.

5.

Acquire and average samples.
a.

Create and configure an AI voltage channel according to Table 2.

b.

Configure the AI voltage channel timing according to Table 3.

c.

Start the task.

d.

Read samples from the NI 9215.

e.

Average the samples for each channel.

f.

Clear the task.

1

Connections when using a calibrator with a guard connection.

2

Connections when using a calibrator with no guard connection.

Table 2.  

NI 9215 AI Voltage Channel Configuration

Input Range

Scaled Units

Terminal Configuration

Min

Max

-10

10

Volts

Differential

Table 3.  

NI 9215 Timing Configuration

Samples Per Channel

Acquisition Mode

Rate (S/s)

10 k

Finite

100 k

+

+

+

+

NI 9215

with BNC

C

a

li

b

r

a

tor

HI

LO

AI0

AI1

AI2

AI

3

+

+

+

+

NI 9215

with BNC

C

a

li

b

r

a

tor

HI

LO

G

ua

rd

AI0

AI1

AI2

AI

3

1

2

Содержание NI 9215

Страница 1: ...irements 2 Test Equipment 3 Test Conditions 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 7 Re Verification 7 Accuracy Under Calibration Con...

Страница 2: ...and Specifications NI 9215 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programming...

Страница 3: ...thermal offsets Maintain an ambient temperature of 23 C 5 C The NI 9215 temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 min...

Страница 4: ...ion Complete the following procedure to determine the As Found status of the NI 9215 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9215 to the calibrator Refer to Figure 1 for a connectio...

Страница 5: ...l according to Table 2 b Configure the AI voltage channel timing according to Table 3 c Start the task d Read samples from the NI 9215 e Average the samples for each channel f Clear the task 1 Connect...

Страница 6: ...ust the NI 9215 a Initialize a calibration session on the NI 9215 The default password is NI b Input the external temperature in degrees Celsius c Call the NI 9215 get C Series adjustment points funct...

Страница 7: ...n Repeat the Verification section to determine the As Left status of the device Note If any test fails Re Verification after performing an adjustment verify that you have met the Test Conditions befor...

Страница 8: ...s in the NI DAQmx Readme Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes E...