manualshive.com logo in svg
background image

6

|

ni.com

|

NI 9215 Calibration Procedure

6.

Compare the average sample for each channel to the test limits in Table 4.

7.

Repeat steps 4 through 6 for each Test Point in Table 4.

8.

Set the calibrator to Standby mode (STBY).

9.

Disconnect the NI 9215 from the calibrator.

Adjustment

The following performance adjustment procedure describes the sequence of operation required 
to adjust the NI 9215.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the NI 9215.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9215 to the calibrator. Refer to Figure 1 for a connection diagram for the 
NI 9215 with screw and refer to Figure 2 for a connection diagram for the NI 9215 with 
BNC.

3.

Adjust the NI 9215.
a.

Initialize a calibration session on the NI 9215. The default password is 

NI

.

b.

Input the external temperature in degrees Celsius.

c.

Call the NI 9215 get C Series adjustment points function to obtain an array of 
recommended calibration voltages for the NI 9215.

d.

Set the calibrator to a reference value determined by the array of recommended 
calibration voltages.

e.

Set the calibrator to Operate mode (OPR).

f.

Call and configure the NI 9215 adjustment function according to Table 5.

g.

Set the calibrator to Standby mode (STBY).

Table 4.  

NI 9215 Test Limits

Range (V)

Test Point

1-Year Limits

Min

Max

Location

Value (V)

Lower Limit (V)

Upper Limit (V)

-10.000

10.000

Max

9.500000

9.494700

9.505300

-10.000

10.000

Mid

0.000000

-0.001500

0.001500

-10.000

10.000

Min

-9.500000

-9.505300

-9.494700

Table 5.  

Adjustment Configuration

Physical Channel

Reference Value

cDAQMod8/ai

x

The reference value from the array of adjustment points

Содержание NI 9215

Страница 1: ...irements 2 Test Equipment 3 Test Conditions 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 7 Re Verification 7 Accuracy Under Calibration Con...

Страница 2: ...and Specifications NI 9215 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programming...

Страница 3: ...thermal offsets Maintain an ambient temperature of 23 C 5 C The NI 9215 temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 min...

Страница 4: ...ion Complete the following procedure to determine the As Found status of the NI 9215 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9215 to the calibrator Refer to Figure 1 for a connectio...

Страница 5: ...l according to Table 2 b Configure the AI voltage channel timing according to Table 3 c Start the task d Read samples from the NI 9215 e Average the samples for each channel f Clear the task 1 Connect...

Страница 6: ...ust the NI 9215 a Initialize a calibration session on the NI 9215 The default password is NI b Input the external temperature in degrees Celsius c Call the NI 9215 get C Series adjustment points funct...

Страница 7: ...n Repeat the Verification section to determine the As Left status of the device Note If any test fails Re Verification after performing an adjustment verify that you have met the Test Conditions befor...

Страница 8: ...s in the NI DAQmx Readme Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes E...

Отзывы: