background image

NI 9215 Calibration Procedure

|

© National Instruments

|

3

Test Equipment

Table 1 lists the equipment required for the performance verification and adjustment procedures. 
If the recommended equipment is not available, select a substitute using the requirements listed 
in Table 1.

Test Conditions

The following setup and environmental conditions are required to ensure the NI 9215 meets 
published specifications.

Keep connections to the NI 9215 as short as possible. Long cables and wires act as 
antennae, picking up extra noise that can affect measurements.

Verify that all connections to the NI 9215 are secure.

Use shielded copper wire for all cable connections to the NI 9215. Use twisted-pairs wire 
to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ±5 °C. The NI 9215 temperature will be greater 
than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 10 minutes to ensure that the NI 9215 measurement 
circuitry is at a stable operating temperature.

Initial Setup

Complete the following steps to set up the NI 9215.
1.

Install NI-DAQmx.

2.

Make sure the NI cDAQ-9178 power source is not connected.

3.

Install the module in slot 8 of the NI cDAQ-9178 chassis. Leave slots 1 through 7 of the 
NI cDAQ-9178 chassis empty.

4.

Connect the NI cDAQ-9178 chassis to your host computer.

5.

Connect the power source to the NI cDAQ-9178 chassis.

Table 1.  

Recommended Equipment

Equipment

Recommended 

Model

Requirements

Calibrator

Fluke 5700A

Use a high-precision voltage source with an 
accuracy 

15 ppm and an output impedance of 

50

Ω

.

Chassis

NI cDAQ-9178

Screw-Terminal 
Connection 
Accessory

NI 9932

Содержание NI 9215

Страница 1: ...irements 2 Test Equipment 3 Test Conditions 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 7 Re Verification 7 Accuracy Under Calibration Con...

Страница 2: ...and Specifications NI 9215 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programming...

Страница 3: ...thermal offsets Maintain an ambient temperature of 23 C 5 C The NI 9215 temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 min...

Страница 4: ...ion Complete the following procedure to determine the As Found status of the NI 9215 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9215 to the calibrator Refer to Figure 1 for a connectio...

Страница 5: ...l according to Table 2 b Configure the AI voltage channel timing according to Table 3 c Start the task d Read samples from the NI 9215 e Average the samples for each channel f Clear the task 1 Connect...

Страница 6: ...ust the NI 9215 a Initialize a calibration session on the NI 9215 The default password is NI b Input the external temperature in degrees Celsius c Call the NI 9215 get C Series adjustment points funct...

Страница 7: ...n Repeat the Verification section to determine the As Left status of the device Note If any test fails Re Verification after performing an adjustment verify that you have met the Test Conditions befor...

Страница 8: ...s in the NI DAQmx Readme Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes E...

Отзывы: