NI 9215 Calibration Procedure

Documentation

Consult the following documents for information about the NI 9215, NI-DAQmx, and your 
application software. All documents are available on 

ni.com

 and help files install with the 

software.

NI cDAQ-9174/9178 Quick Start

NI-DAQmx installation and hardware setup

NI 9215 Operating Instructions and Specifications

NI 9215 specific information, specifications, and calibration interval

NI-DAQmx Readme

Operating system and application software support in NI-DAQmx

LabVIEW Help

LabVIEW programming concepts and reference information about NI-DAQmx VIs and 
functions

NI-DAQmx C Reference Help

Reference information for NI-DAQmx C functions and NI-DAQmx C properties

NI-DAQmx .NET Help Support for Visual Studio

Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET properties, key 
concepts, and a C enum to .NET enum mapping table

Содержание NI 9215

Страница 1: ...irements 2 Test Equipment 3 Test Conditions 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 7 Re Verification 7 Accuracy Under Calibration Con...

Страница 2: ...and Specifications NI 9215 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programming...

Страница 3: ...thermal offsets Maintain an ambient temperature of 23 C 5 C The NI 9215 temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 min...

Страница 4: ...ion Complete the following procedure to determine the As Found status of the NI 9215 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9215 to the calibrator Refer to Figure 1 for a connectio...

Страница 5: ...l according to Table 2 b Configure the AI voltage channel timing according to Table 3 c Start the task d Read samples from the NI 9215 e Average the samples for each channel f Clear the task 1 Connect...

Страница 6: ...ust the NI 9215 a Initialize a calibration session on the NI 9215 The default password is NI b Input the external temperature in degrees Celsius c Call the NI 9215 get C Series adjustment points funct...

Страница 7: ...n Repeat the Verification section to determine the As Left status of the device Note If any test fails Re Verification after performing an adjustment verify that you have met the Test Conditions befor...

Страница 8: ...s in the NI DAQmx Readme Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes E...