NI 9215 Calibration Procedure

6.

Launch Measurement & Automation Explorer (MAX).

7.

Right-click the device name and select 

Self-Test

 to ensure that the module is working 

properly.

Verification

The following performance verification procedure describes the sequence of operation and test 
points required to verify the NI 9215 and assumes that adequate traceable uncertainties are 
available for the calibration references.

Accuracy Verification

Complete the following procedure to determine the As-Found status of the NI 9215.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9215 to the calibrator. Refer to Figure 1 for a connection diagram for the 
NI 9215 with screw and refer to Figure 2 for a connection diagram for the NI 9215 with 
BNC.

Figure 1.  

Accuracy Connections for the NI 9215 with Screw Terminal

1

Connections when using a calibrator with a guard connection.

2

Connections when using a calibrator with no guard connection.

NI 9215 with

s

crew termin

a

l

C

a

li

b

r

a

tor

HI

LO

AI0

+

AI1

+

AI2

+

AI

3

+

COM

NI 9215 with

s

crew termin

a

l

C

a

li

b

r

a

tor

HI

LO

G

ua

rd

AI0

+

AI1

+

AI2

+

AI

3

+

COM

1

2

Содержание NI 9215

Страница 1: ...irements 2 Test Equipment 3 Test Conditions 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 7 Re Verification 7 Accuracy Under Calibration Con...

Страница 2: ...and Specifications NI 9215 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programming...

Страница 3: ...thermal offsets Maintain an ambient temperature of 23 C 5 C The NI 9215 temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 min...

Страница 4: ...ion Complete the following procedure to determine the As Found status of the NI 9215 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9215 to the calibrator Refer to Figure 1 for a connectio...

Страница 5: ...l according to Table 2 b Configure the AI voltage channel timing according to Table 3 c Start the task d Read samples from the NI 9215 e Average the samples for each channel f Clear the task 1 Connect...

Страница 6: ...ust the NI 9215 a Initialize a calibration session on the NI 9215 The default password is NI b Input the external temperature in degrees Celsius c Call the NI 9215 get C Series adjustment points funct...

Страница 7: ...n Repeat the Verification section to determine the As Left status of the device Note If any test fails Re Verification after performing an adjustment verify that you have met the Test Conditions befor...

Страница 8: ...s in the NI DAQmx Readme Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes E...

Отзывы: