only after the adapter module has been powered on by the NI FlexRIO FPGA
module.
Caution
Connections that exceed any of the maximum ratings of any connector on
the NI 5791R can damage the device and the chassis. NI is not liable for any damage
resulting from such connections.
Table 2. NI 5791 Front Panel Connectors
Device Front Panel Connector
Signal Description
AUX
I/O
NI 5791
200MHz - 4.4 GHz
RF Transceiver
RX IN
LO OUT
CLK OUT
LO IN
TX OUT
CLK IN
INSTRUMENTS
NATIONAL
RX IN
Receive channel input, +20 dBm maximum
LO OUT
Local oscillator output, +12 dBm maximum, +0 dBm
nominal
CLK IN
Reference Clock input, 50 Ω single-ended, +20 dBm
maximum
CLK OUT Exported clock output, DC-coupled, 0 V to 2 V
LO IN
Local oscillator input, +20 dBm maximum
TX OUT
Transmit channel output, +20 dBm maximum
AUX I/O
Refer to Table 3 for the signal list and descriptions.
Related Information
NI 5791 Specifications
on page 21
NI 5791R User Manual and Specifications | © National Instruments | 7