Caution
Disconnect external power to the MON-10467 (AC) and use the external
circuit breaker box to completely remove mains power before opening the device for
service.
Attention
Débranchez l'alimentation externe du MON-10467 (AC) et utilisez le
boîtier de disjoncteur externe pour couper complètement l'alimentation secteur avant
d'ouvrir l'appareil pour l'entretien.
Caution
Remove power to the MON-10467 (DC) before opening it for service.
Attention
Mettez le MON-10467 (DC) hors tension avant de l'ouvrir pour
l'entretien.
Notice
Do not put a load on the device door while it is open.
Figure 12. Puncturing and Inserting the Desiccant Pack
Puncture Here
Complete the following steps.
1.
If you are using the MON-10467 (AC) or MON-10467 (DC), confirm that the power to
the device has been turned off.
2.
Loosen the four captive screws on the door of the device.
3.
Open the door of the device slowly.
4.
Activate the desiccant pack by piercing the Mylar outer bag with the provided 1/16 in.
punch. See the figure above for where to puncture the desiccant pack.
MON-10467 User Guide
|
© National Instruments
|
27