Integrated Electronic Piezoelectric (IEPE) Sensors
The MON-10467 can provide an IEPE excitation current for each channel to measure IEPE
sensors. You can configure IEPE excitation current per channel.
Typical IEPE sensors have a case that is electrically isolated from the IEPE electronics.
Therefore, connecting this type of sensor to the MON-10467 results in a floating connection.
Typical IEPE sensors used in industrial machine condition monitoring applications may also
be supplied with a cable shield drain wire that is floating at the sensor end. In this case, the
drain wire should be connected to the channel chassis ground terminal of the MON-10467.
IEPE Fault Conditions
When IEPE is enabled, the MON-10467 monitors the input voltage to determine if there is a
connection issue with the attached signal source. Common connection issues that can occur
include open and short circuits. An open circuit occurs if there is a disconnect between the
signal source and the MON-10467, resulting in the input voltage rising too high. A short
circuit occurs if the MON-10467 AI+ to AI- terminals have been shorted. For example, a
sensor fault could cause a short circuit. If an open or short event occurs, the hardware will
continue to perform the configured acquisition and return the acquisition to InsightCM, but
will set a fault indicator flag advising the user that the returned data is invalid.
Connecting Power and Ground
MON-10467 (AC) Connecting Power and Ground
Hardware needed:
•
PG16 gland (supplied by NI) or 0.5 in. conduit fitting
•
Power cable with a smooth, round jacket and individual conductors up to 12 AWG that
comply with your local electrical code
Tools needed:
•
2.5 mm flat-blade screwdriver (your own or NI part number 786911-01)
•
27 mm deep socket
•
30 mm socket or wrench
•
24 mm socket or wrench
•
(Optional) 24 mm gland dome nut tightening tool (SEALCON part number S-2400-WR)
Caution
Disconnect external power to the MON-10467 (AC) and use the external
circuit breaker box to completely remove mains power before opening the device for
service.
Attention
Débranchez l'alimentation externe du MON-10467 (AC) et utilisez le
boîtier de disjoncteur externe pour couper complètement l'alimentation secteur avant
d'ouvrir l'appareil pour l'entretien.
Complete the following steps.
1.
Confirm that the external power to the MON-10467 (AC) has been turned off.
2.
Install the gland on the device enclosure.
22
|
ni.com
|
MON-10467 User Guide