ECUTS-16000 Environmental, Physical, and Interface Requirements
Ventilation
Interface
7
4
2
3
3
5
1
1
2
3
4
5
Environment
Physical
Ensure the location has stable
environmental conditions to reduce
variability in performance.
Ambient Temperature
5 °C to 40 °C (41 °F to 104 °F)
Relative Humidity
20% to 80%, noncondensing
Pollution Degree
2
Indoor use only.
Ensure the facility has cooling capacity
in excess of the maximum power
consumption of the ECUTS.
Ensure the location can accommodate
the size and weight of the ECUTS.
Test Rack Dimensions
584 mm x 853 mm x 1440 mm
(23 in. x 33.6 in. x 56.5 in.)
(Optional) Monitor Maximum Height
2080 mm (81.9 in.)
Depth Including Optional 9025TR
Platform
1650 mm (65 in.)
Maximum Weight
240 kg (529 lb)
Do not obstruct the required
clearances to ensure proper air flow.
Obstruction thresholds are measured
from the test rack exterior.
Top
760 mm (30 in.)
Front and Back
300 mm (12 in.)
Provide AC power connection to
facility power source.
Voltage
200 VAC to 240 VAC, single phase
Maximum Current
16 A
Frequency
50/60 Hz
Maximum Branch Circuit
20 A
AC Inlet Type
IEC 60320 C20
Provide connection to facility
protective earth ground.
6
(Optional) Provide a clean, dry
compressed air connection.
Pneumatic Air Pressure Rating
100 PSI maximum
7
Provide network connection.
44
|
ni.com
|
ECUTS-16000/16001 User Manual