30 m (100 ft). The length of any cable connected to the USB host ports must be no
longer than 5 m (16 ft).
Caution
The USB device port is intended only for use in device configuration,
application deployment, debug, and maintenance.
Caution
To ensure the specified EMC performance, product installation requires
either special considerations or user-installed, add-on devices. Refer to the following
sections for more information.
Related Information
Connecting the cRIO-9037 to Ground
Connecting the cRIO-9037 to Power
on page 9
Special Conditions for Marine Applications
Some products are Lloyd’s Register (LR) Type Approved for marine (shipboard) applications.
To verify Lloyd’s Register certification for a product, visit
and search for
the LR certificate, or look for the Lloyd’s Register mark on the product.
Caution
In order to meet the EMC requirements for marine applications, install the
product in a shielded enclosure with shielded and/or filtered power and input/output
ports. In addition, take precautions when designing, selecting, and installing
measurement probes and cables to ensure that the desired EMC performance is
attained.
Preparing the Environment
Ensure that the environment in which you are using the cRIO-9037 meets the following
specifications.
Operating temperature
(IEC 60068-2-1, IEC 60068-2-2)
-20 °C to 55 °C
Operating humidity (IEC 60068-2-78)
10% RH to 90% RH, noncondensing
Pollution Degree
2
Maximum altitude
5,000 m
Indoor use only.
Note
Refer to the device specifications on
specifications.
4
|
ni.com
|
NI cRIO-9037 Getting Started Guide