National Instruments Corporation
23
NI 5762R User Guide and Specifications
Note
Clean the device with a soft, non-metallic brush. Make sure that the device is completely dry
and free from contaminants before returning it to service.
Shock and Vibration
Operational shock ..................................................30 g peak, half-sine, 11 ms pulse,
tested in accordance with IEC-60068-2-27.
Test profile developed in accordance with
MIL-PRF-28800F.
Random vibration
Operating .......................................................5 Hz to 500 Hz, 0.3 g
rms
Nonoperating .................................................5 Hz to 500 Hz, 2.4 g
rms
,
tested in accordance with IEC-60068-2-64.
Nonoperating test profile exceeds the requirements
of MIL-PRF-28800F, Class 3.
Safety
This product meets the requirements of the following standards of safety for electrical equipment for
measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for
measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note
For EMC declarations and certifications, refer to the
section of
this document.
CE Compliance
This product meets the essential requirements of applicable European Directives as follows:
•
2006/95/EC; Low-Voltage Directive (safety)
•
2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
To obtain product certifications and the Declaration of Conformity for this product, visit
ni.com/
certification
, search by model number or product line, and click the appropriate link in the
Certification column.