NI 5762R User Guide and Specifications
22
ni.com
Maximum toggle frequency...................................6.6 MHz
+5 V maximum power ...........................................10 mA
+5 V voltage tolerance ...........................................4.2 V to 5 V
Configuration EEPROM Map
Caution
Only
write to
User Space
. Writing to any other offset may cause the NI 5762 to stop
functioning.
Power
Total power, typical operation................................5.3 W
Physical
Dimensions ............................................................12.9
×
2.0
×
12.1 cm
(5.1
×
0.8
×
4.7 in.)
Weight ....................................................................332 g (11.7 oz)
Front panel connectors...........................................SMA and HDMI
Environmental
The NI 5762 is intended for indoor use only.
Operating environment
1
.........................................0
°
C to 55
°
C,
tested in accordance with IEC-60068-2-1 and
IEC-60068-2-2.
Relative humidity range .........................................10% to 90%, noncondensing,
tested in accordance with IEC-60068-2-56.
Altitude ..................................................................2,000 m at 25
°
C ambient temperature.
Pollution Degree ....................................................2
Storage environment
Ambient temperature range ...........................–20
°
C to 70
°
C,
tested in accordance with IEC-60068-2-1 and
IEC-60068-2-2.
Relative humidity range .................................5% to 95%, noncondensing,
tested in accordance with IEC-60068-2-56.
Byte Address
Size (Bytes)
Field Name
0x0
2
Vendor ID
0x2
2
Product ID
0x4
4
Serial Number
0x8
116
Reserved
0x7C
132
User Space
1
For PXI/PXI Express chassis configurations that group NI FlexRIO adapter modules in three or more contiguous slots,
National Instruments recommends limiting the ambient operating temperature to less than 50
°