16
- Hold the tip wire with the tweezers behind the freshly cut tip.
Important!
• Do not touch the end of the tip with anything.
• Take care that the tip wire is straight.
• Do not twist the tip clamp in the open part of the scan head nor lift it too
high.
The procedure for mounting the tip under the tip clamp depends on the
type of clamp on your scan head:
Single clamp
(see figure Mounting the tip under the single clamp):
a)
b)
2.
1.
Mounting the tip under the single clamp
1. Put the tip wire on the tip holder parallel to the groove in the tip holder,
so that it crosses below the tip clamp (figure a).
2. Move the tip wire sideways until it is in the groove in the tip holder.
(Figure b)
The freshly cut tip should be securely held under the clamp and extend
about 2-3mm beyond the tip holder. The tip is now installed.