NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
78
Note
: If samples with a different wavelength range are selected, the graphs are shown on
full scan range of 200–900 nm (N50: 200-650 nm).
6. Baseline correction is set to off as default. Enabling the baseline corrections shows a list
with different wavelength options: 377 nm, 604 nm, 650 nm, 770 nm (N/A N50) and 823
nm (N/A N50). Option to enter any wavelength between 200 nm and 900 nm (N50: 650
nm)
7. Option to smooth the graph with different boxcars. Options: Off, 1 = boxcar 11 (default), 2
= boxcar 21 and 3 = boxcar 61
8. Option to set/calculate a dilution factor for manual diluted samples.
9. Apply the blank ddH
2
0 or buffer to the illuminated sample window for the
reference measurement and select blank to initiate the reading.
Note:
The illumination of the sample window can be switched off in the
preferences.