NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
46
Note:
A new Blank is recommended when changing the dilution lid.
5. If it is desired to heat the sample to 37°C use the toggle switch to turn on
cell holder heater. When the cuvette holder has reached 37°C the toggle
color changes to green.
Note
: Only available for cuvette applications (NP80 and C40).
6. The background correction is enabled at 320 nm as a default. It is possible to disable the
background correction with the toggle switch. Selection options are 320 nm, 340 nm. It is
also possible to add any wavelength in the range of 220 – 350 nm
7.
8. Air bubble recognition is disabled as a default. Set to on it detects air
bubbles, lint residues and bad sample conditions of the sample.
Note
: Lint residues and bad sample conditions are detected even if the
air bubble recognition is set to off.
9. For dye labelled samples tick the dye labels in the list, which should be used for result
calculation.
Note:
If the used dye is not available in the list, please go to preferences and add a custom
dye to the dye list.
There is an option for dye correction which can be enabled/disabled with a toggle switch.
Note:
Dye correction is only available for single dye selection.