4. Determine the value for the
minimum cell size
and the
maximum cell size
touching the control button ( ) to adjust
minutely. The set-value will be immediately reflected in the
Counting result
.
·
Minimum cell size
is used to determine the low range of cell size within the
measurement. The algorithm first identifies all objects and calculates the average
size. From the percent of average size setting, the algorithm calculates the smallest
object size to include in the final measurement. Adjusting the number up, increases
inclusiveness thereby decreasing the lower cell size range.
·
Maximum cell size
is used to determine the high range of cell size within the
measurement. The algorithm first identifies all objects, then calculates the average
size. From the percent of average size setting, the algorithm calculates the largest
object size to include in the final measurement.
5. Cell counting result will be shown with cell size distribution
graph. If you are not satisfied with the result, then try again after
adjusting minimum and maximum cell size.
Operation
Counting
Bright counting
(Stained by trypan blue)
JuLI
TM
FL, Fluorescence Cell History Recorder ©2013 NanoEnTek Inc.
37