ADAM - rWBC ©2013 NanoEnTek Inc.
17
Starts the counting process by the Start button input.
Once the focusing process finished after calibration step,
ADAM-rWBC will skip the focusing process for the repeated experiments.
Captures the images from the valid r-Slide measuring area and ana-
lyzes the images in memory and performs the counting process.
The number of total leukocytes will be automatically calculated by the
ADAM-rWBC software.
Start input
Result
analysis
General operation
Number of total leucocytes
1.71/µL
• Error message during result analysis
[Bad sample!]
In the event that 1 cell image contains of more than 100 pixels includ-
ing dust or the number of error frame is more than 10 frames, this error
message will be displayed on the LCD screen. The error frame occurs
when the number of cells is more than 50 in one frame. Check the
sample contains dust or cell clumping and retest.