ADAM - rWBC ©2013 NanoEnTek Inc.
11
[ System init. ] If a r-Slide is detected during initial booting, this error
message will be shown.
(Trouble shooting) 1. Remove the r-Slide 2. Power off/on 3. Redo
initial booting again.
CAUTION:
When you turn off the device, please remove the r-Slide
from the ADAM-rWBC.
This is the cause of malfunction.
If this message continues to display when no r-Slide is inserted,
contact your local distributor or [email protected].
* Error Messages during booting
[System State] It appears when booting not working properly. Turn off
main power and re-start device.
* Power switch is located back side of instrument.
If this message still appears after re-start, contact your local distributor
or
System installation
Power on and
initial display
Error code
0x00000C00
0x00007000
0x00008000
0x06000000
Cause
Failure of X-axis sensor
Failure of Y-axis sensor
Failure of Z-axis sensor
Failure of locking module sensor