CAUTION:
At low temperatures (≤10 °C), allow the device to warm up
for 10 minutes at ambient temperature before use.
To ensure correct operation and stable performance,
install the ADAM-rWBC in a location which meets the following conditions:
• Use at room temperature.
• Do not use in cold rooms.
• Not exposed to direct sunlight.
• Not subject to direct or continuous vibration.
• Not subject to intense magnetic or electromagnetic fields.
• Not installed in high humidity environments.
• Area free from corrosive gases or other corrosive substances.
• Area with very little dust or other airborne particles.
• Area with a 10 cm minimum space around the instrument for
proper air flow.
• Do not place anything on top of ADAM-rWBC device.
System installation
Environmental
requirements
Power on and
initial display
1) Check the connection of the main device power cord.
2) Make sure that the main power switch is in the “I” (ON) position.
(On the rear side of the main device.)
• When you turn on the ADAM-rWBC, it will go through self diagnostic
tests. If you get an error message, please contact your local distributor
or [email protected]. If boot up is successful and no errors are
detected, the home screens will be displayed as follow.