14
1. Leave the r-Kit at room temperature for 10 minutes before use.
2. Carefully dispense 100 uL of well-mixed RBC or platelet sample into
a clean test tube.
3. Add 400 µL of reagent into the tube.
4. Mix the tube well (100 µL 400 µL reagent).
5. Load 100 µL mixed sample/reagent onto a r-Slide.
6. Wait 4~7 minutes for sample settling.
7. Insert r-Slide into ADAM instrument and press ' ' to start.
8. After approximately 3 minutes, the result, calculated as WBC per µL,
will be displayed automatically.
General operation
Sample
preparation
and testing
1. Mix sample with
r-Solution & Load sample
(100μL)
2. Insert r-Slide
3. Result
r-Solution
400 µL
100 µL