22
Introduction
QC
QC Mode
The QC Mode uses QC Slide (optional) to check equipment QC status by
date at a glance.
[ Activation of QC mode ]
To activate ‘QC’ mode, an activation code must be entered.
1. Select ‘QC’ tab from top menu.
2. Enter lot no. and activation code. Then, click ‘APPLY’ button.
Slide lot no. and activation code can be found on the plastic package
label. See below for details.
Control buttons
Description
①
Data Type
QC result graph unit (Years, Months, Days,No.[Index])
②
Show Data
Displays the date of QC progress on the graph
③
QC Result Graph
Graph left/right (QC date, number) movement button
④
Arrow Button
Graph left/right (QC date, number) movement button
⑤
QC Slide
QC Slide Lot. information and editing (create, edit,
delete) functions
⑥
QC Result
QC result information and editing (whether or not
graph is displayed, deleted) function
⑦
SAVE
Saves the QC Result Report to USB
⑧
Sends the QC Result Report to e-mail
QC Mode must use the QC Slide (optional), and the result without using
the QC Slide is unreliable.
!
WARNING
A unique activation code is given for each instrument, and its authenticity
can be checked by registering activation code.
!
NOTE
①
②
③
④
⑦
⑧
⑤
⑥