12
Introduction
Power on and
Initial Display
Getting Started
Environmental
requirements
To ensure correct operation and stable performance,
install the ADAM-CellT in a location which meets the following conditions:
1. Use at room temperature between 20 and 35 °C
• Not recommended for cold room use (≤ 4 °C).
2. Do not expose the device to direct sunlight.
3. Do not subject the device to direct or continuous vibration.
4. Do not subject the device to intense magnetic or electromagnetic fields.
5. Do not install the device in high-humidity environment.
6. Location of device should be free from corrosive gases or other corrosive
substances.
7. Ensure minimal contact with dust or other airborne particles.
8. Allow a 10 cm (4 inches) minimum space around the device for proper
airflow.
9. Do not place any objects on the device.
1. Check the connection of ADAM-CellT and power cord.
2. Press the power button for 2~3 seconds.
If you get an error message,
please contact your local distributor or [email protected].
If booting is successful and no errors are detected,
the home screens will be displayed as below.
At low temperature (≤10 °C), allow the device to warm up for 10 minutes at
ambient temperature before use.
!
CAUTION
•Do not tilt the device too much in the forward when connecting the power cord.
•Do not move the device after connecting power cord.
- When you connect the power cord to ADAM-CellT even without power on the device,
it will go through self diagnostic tests.
!
CAUTION