20
Zetasinter System
®
7.
Vacuum pump & flowmeter
Vacuum pump
①
Unscrew the blue vacuum pump cap and inject the appropriate amount of oil. Put back the vacuum
pump cap without (without the small black cover).
②
Plug the vacuum pump into the socket on the side of the furnace body.
③
Connect the vacuum pump to the outlet flange using the Ø10mm flexible air hose.
1)
2)
.
Flowmeter
①
Install the flowmeter on its holder.
②
Connect its outlet (upwards) to the inlet flange with the Ø6mm flexible air hose.
③
Connect its inlet (downwards) to the 0.2 bar (0.02MPa) gas supply (1/8 NPT F).
1)
2)
3)
Use: Flushing
-
Close the valves and the flowmeter.
-
Turn on the pump.
-
Open the outlet valve.
-
Wait till the tube pressure is under -0.9 bar (-0.09 MPa).
-
Close the outlet valve.
Содержание Zetasinter
Страница 12: ...12 Zetasinter System Specific settings Time Setting Heating programs saving ...
Страница 16: ...16 Zetasinter System Install the right flange inlet flange 1 2 3 4 5 Install the left flange outlet flange ...
Страница 26: ...26 Zetasinter System 11 Electrical schematic diagram 220V Electrical schematic diagram ...