Mytrex MXD3G User’s Manual 
Rev 0 July 29, 2013 

Page 14 of 21 

3.

 

Press and release the base unit HELP button a second time.  The base unit will announce, 

“Program Personal Help Buttons Group Two.”

   

4.

 

Program PHB codes into Group 2 by following steps 3 through 6 in the instructions for 
programming PHBs into Group 1.  

5.

 

If PHB codes are to be programmed into Group 3, press the HELP button again. The base unit 
will announce, 

“Program Personal Help Buttons Group Three.”

   

6.

 

Program PHB codes into Group 3 by following steps 3 through 6 in the instructions for 
programming PHBs into Group 1.  

7.

 

If PHB codes are to be programmed into Group 4, press the HELP button again. The base unit 
will announce, 

“Program Personal Help Buttons Group Four.”

   

8.

 

Program PHB codes into Group 4 by following steps 3 through 6 in the instructions for 
programming PHBs into Group 1.  

9.

 

After all PHBs are programmed into the desired groups, press the RESET button one time.  The 
MXD3G will announce, “

[number] Bars, System Ready.”

   This completes the programming 

sequence, and each of the PHBs recorded in Group 1 should now be recognized by the unit. 

10.

 

Test all of the PHBs ONE AT A TIME by pressing each button and allowing the unit to call the 
response center.  Wait to start the next PHB test until the unit has disconnected from the 
previous test call.  Check carefully with the response center attendant to verify that each PHB 
has been programmed into the correct PHB group.  

11.

 

If each test call is successful, the PHB is ready for service.  Reprogram any PHBs into the desired 
group that do not activate the base station when pushed.   

Programming Medication Dispensers 

Note: Medication dispensers may not be supported by your service provider.  Contact your service 
provider to determine if this option is available.  

 
To program a Medication Dispenser Code into the MXD:  

The MXD3G unit can be programmed to 

recognize up to two Mytrex approved medication dispensers that have been equipped for monitoring 
with the MXD3G.     
 
Preparing the Medication Dispenser:   Perform the following steps for each medication dispenser that 
the MXD3G will be programmed to recognize:  

1.

 

Connect the medication dispenser to its power cord and plug the power block into the wall.  

2.

 

Make sure that the medication dispenser is turned on 

3.

 

Unlock and remove the medication dispenser’s top cover 

4.

 

Remove the medication dispenser’s circular pill tray from the dispenser.  

5.

 

Locate the transmitter circuit board underneath the circular pill tray, between the dispenser 
center mechanism cover and the outer wall of the dispenser case.  

6.

 

Locate the small test button on the transmitter circuit board.  It is a silver round button 
approximately 3/16 inches in diameter.   

 

 

 

Содержание MXD3G

Страница 1: ...MXD3G User Manual Revision 0 July 29 2013 Manufactured by Mytrex Inc 10321 South Beckstead Lane South Jordan UT 84095 801 571 4121 Fax 801 571 4606 Toll Free 800 688 9576 Fax 877 571 4606 ...

Страница 2: ...p Buttons PHB and Other Devices Overview 12 MXD3G PHB and Other Device Groups 12 Programming PHBs into Groups 13 Programming Medication Dispensers 14 Programming Smoke Detectors Supervised Devices and CO Detectors 15 Troubleshooting Use of Status Lights 16 Troubleshooting Use of MXD3G Spoken Messages 17 Notifications 19 Notice to End User Wireless Telephone Reliability Considerations 20 When Using...

Страница 3: ...mily friends or emergency services based on the assistance needed Even if you cannot hear or speak to the attendant help will be summoned for you The MXD3G unit is fast and simple to install Your service provider s representative may connect the MXD3G unit during an installation visit you may elect to do it yourself or you may ask a friend or relative to install it for you To ensure that you alway...

Страница 4: ...3G base unit do not place the MXD3G base unit near any liquids Do not spray or wipe cleaners directly on the MXD3G base unit Extensive moisture can cause damage to the MXD3G unit Do not abuse the power cord Never carry the unit by the power cord or yank the cord to disconnect it from an electrical outlet Instead grasp the plug and pull to disconnect ...

Страница 5: ...ure 1 Base Unit Top View Figure 2 Base Unit Back View Base Unit Help Button Speaker Status Light Microphone Power Block and Cord not removable from base unit Cellular Antenna Test Learn T L Button RESET Button Programming Jack for service provider use only Cellular Antenna ...

Страница 6: ...back of the MXD3G These two buttons and the base unit HELP button on the top of the unit are used to conduct range tests turn off the MXD program the MXD3G to recognize personal help buttons PHBs and for other uses See pages 10 16 Two Way Voice Communications After connecting to the response center your MXD3G unit shifts to a voice communication mode The MXD3G usually operates in duplex mode where...

Страница 7: ...l Features Automatic Testing Your MXD3G has the capability of being programmed to carry out automatic tests Your service provider decides whether such tests will be conducted their frequency and the type of tests to be done as part of the service that the provider offers These tests may include automatic voice tests during which the MXD3G sends a test signal to the response center and then connect...

Страница 8: ...tendant Even if you cannot communicate with your response center attendants are trained to treat every received alarm as a call for help until they are sure you are okay Is my MXD3G compatible with all cellular networks The MXD3G is only designed to operate with the AT T 3G Cellular Network It is not designed to work with any other cell network or with any land line telephone service Please contac...

Страница 9: ...base unit in the locations where you plan to use it in order to determine that solid repeatable cell service is present at all times Setting Up Your MXD3G Your MXD3G allows you to request assistance 24 hours per day seven days per week by simply pressing your Personal Help Button Figure 4 Box Contents 1 MXD3G All in One Cellular PERS device 3 Personal Help Button PHB 2 7 inch Cellular Antenna 4 Qu...

Страница 10: ...t again announces Cell Module not activated please contact your service provider for assistance 3 If the MXD again announces Insufficient Cell Service please relocate the base unit to another place in the residence and plug it into a nearby AC outlet that is not controlled by a switch Wait 30 seconds and then press the RESET button If the unit continues to announce Insufficient Cell Service try an...

Страница 11: ...number bars System Ready Testing the MXD3G unit Note The manufacturer recommends that the MXD3G be tested at least every 30 days Most response centers welcome regular testing Attendants are available 24 hours a day to help Please check with your provider concerning their testing policy 1 Press the PHB and wait for the response center attendant to speak to you through the MXD3G 2 When the attendant...

Страница 12: ... to verify that they are programmed correctly and are reporting to the response center under the correct group The following paragraphs describe each of the above steps in more detail MXD3G PHB and Other Device Groups Device codes are stored in one or more groups for each type of device The following groups are available for each type of device Type of Device Number of Groups Number of Devices tha...

Страница 13: ...on is released This is the signal that the unit has recognized and stored the PHB code 4 After at least one tone is received stop pressing the PHB 5 If more than one PHB is to be programmed into Group 1 press and hold the second PHB until the unit sounds a the tone sequence of one second tone followed by two seconds of silence 6 Repeat the process for up to a total of four PHBS in Group 1 7 The MX...

Страница 14: ...B test until the unit has disconnected from the previous test call Check carefully with the response center attendant to verify that each PHB has been programmed into the correct PHB group 11 If each test call is successful the PHB is ready for service Reprogram any PHBs into the desired group that do not activate the base station when pushed Programming Medication Dispensers Note Medication dispe...

Страница 15: ...ars System Ready 9 Initiate a test call for each medication dispenser by pressing and releasing the small button located on the transmitter inside the dispenser The MXD3G should detect the medication dispenser signal and initiate a call to the response center An attendant should answer the call 10 Important Verify with the response center attendant that he she received a medication dispenser signa...

Страница 16: ...dvanced into the desired group category press the HELP button to move from group 1 through 8 8 When the desired devices have been programmed press and release the RESET button to return the unit to normal operation The MXD3G will announce number Bars System Ready 9 Initiate a test call from each programmed device by pressing and releasing the alarm or device test button The MXD3G should detect the...

Страница 17: ...e MXD3G has not been set up with Telular and calls cannot be processed If received during initial installation follow the instructions under Installation Steps page 10 If received after the unit has been installed and tested try pushing the RESET button and listen for a System Ready message If the not activated message persists please contact your service provider Cell module suspended The cell mo...

Страница 18: ...esponse center or has malfunctioned Contact your service provider for further instructions Program Medication Dispenser The unit is ready to be programmed to accept signals from a medication dispenser Follow the instructions as directed on Page 14 Programming mode The unit is in an advanced mode for programming For provider use only Press the RESET button to return the unit to normal operation Pro...

Страница 19: ...rate the equipment Interference Information FCC Rules Part 15 The MXD3G complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Note Your MXD3G has been tested and found to comply with the limits for a...

Страница 20: ...et can affect cell signal connection quality It is possible that your unit will not be able to complete a call because of the lack of cell service even though a connection was made without problems when the unit was first installed and tested A large number of cell phone users trying to use the local network at the same time may overload cell circuits and cause your MXD3G to detect a busy signal i...

Страница 21: ...Before choosing the MXD3G for your medical alarm needs it is extremely important that you press your button to initiate tests of the MXD3G in the locations where you plan to use it most frequently in order to determine that solid repeatable cell service is present at all times NOTICE THE MONITORING CENTER AND THE COMPANIES PROVIDING YOUR MEDICAL ALARM SERVICE ARE NOT LIABLE FOR PROBLEMS RELATED TO...

Отзывы: