background image

 

6

 

 

 

listed above. For example, if the activity of Cl

-

 is ten times higher than nitrate in the sample, Cl

-

 may cause 7% error for the nitrate 

measurement, while CH

3

COO

-

 and SO

4

2- 

in the same ratio may only cause 0.7% and 0.05% error, respectively.    

When testing samples with high interference levels, such as wastewater, chlorinated water, and soil water, the interference ions 
will be counted as nitrate with the coefficient factor and cause measurement errors. Some commonly existing interference ions, 
such as Cl

-

, Br

-

, I

-

, S

2-

, HCO

3

-

 and NO

2

-

 can be reduced by using Nitrate Interference Suppression Buffer (NISB). Add NISB at 1:1 

volume ratio to all sample/standard solutions (e.g. add 5 mL NISB to 5 mL solution) in place of ISA throughout the sensor 
conditioning, calibration and measurement.  

NOTE

: Nitrate Interference Suppression Buffer (NISB) is purchased separately, see ACCESSORIES. 

NOTE

: When using NISB, do NOT add ISA. White precipitate may form in sample solution after mixing with NISB, it will not 

affect the measurement. 

NOTE

: Strong interference ions such as ClO

4

-

 and SCN

-

 require their concentrations to be up to 2000x lower than nitrate in the 

sample to avoid high errors.  

IV. MEASUREMENT VALIDATION METHODS

1.

 

Calibration Check: 

 

a.

 

Choose one standard solution within the testing range, and prepare the solution according to the measurement procedure. 

b.

 

Test readings in the standard solution and the value should be within 10% of the expected value. If not, the measurement 
result  may  not  be  reliable.  The  source  of  error  should  be  identified  and  the  measurement  should  be  redone  after 
correction.  

2.

 

Known Addition Recovery Check:

 

This  validation  method  works  well  for  validating  low  concentration  measurement  (1-60  ppm-N)  by  using  concentrated 
standard solutions to increase the nitrate concentration in the sample by a known value, optimally double the measured value, 
and then compare the value with the increase in Pen readings. For about 10 mL sample, one spike of standard solution with 
the  100 

µ

L  pipette  will  increase  the  sample  nitrate  concentration  by  roughly  1%  of  the  standard  solution  concentration, 

example ~1 ppm increase by one spike of 100 ppm standard solution, ~10 ppm increase by one spike of 1000 ppm standard 
solution.  
    
Spike reference for different sample concentration for 10 mL sample volume (

V

/mL) 

Measured Sample Concentration (

ppm NO

3

-

-N

×

100 

µ

L pipette 

Standard Solution (

ppm NO

3

-

-N

1-3 

100 

3-9 

100 

9-20 

1000 

20-60 

1000 

a.

 

Add 

n

 spikes (volume increase: 

𝑽

𝒂𝒅𝒅

=0.1

×

n/mL) of standard solution (

Cs

/ppm) to the sample solution after measurement 

(

C1

/ppm) and swirl for mixing. 

b.

 

Take the 2

nd

 measurement (

C2

/ppm) and calculate the recovery

 

using the following equation. 

 

𝑅𝑒𝑐𝑜𝑣𝑒𝑟𝑦

%

=

 

𝐶

!

×

100

𝑉

×

𝐶

!

+

𝑉

!""

×

𝐶

!

×

(

1

/

(

1

𝐷

!

)

)

(

𝑉

!""

+

𝑉

)

 

Where  Df  is  the  dilution  factor  when  spiking  unprepared  standard  solutions,  1/51  for  using  ISA  and 

½

  for  using  NISB.  If 

adding prepared standard solutions, Df is 0. 

The recovery should be within 10% of the expected value. If not, the measurement result may not be reliable. The source of 
error should be identified and the measurement should be redone after correction. 

MAINTENANCE 

I. ROUTINE MAINTENANCE  

1.

 

After each use 

ALWAYS

 rinse the Nitrate sensor with clean water (preferably DI, RO, or distilled) and then carefully pat the 

sensor with a clean, soft, lint-free cloth to remove any water drops.  

2.

 

ALWAYS

 replace the protective cap on the Nitrate sensor after each use.  

3.

 

Do not touch the ISE membrane as oil from your finger may contaminate the sensor. 

4.

 

Do not drop, throw, or otherwise strike the PT6 as this may void the warranty. 

5.

 

Do not store the PT6 in a location where the ambient temperatures exceed its operating/storage temperature limits. 

Содержание Ultrapen PT6

Страница 1: ...he sensor submerged 5 In HOLD mode Real time readings are displayed until the LED turns solid ON The display holds the final reading briefly then the PT6 turns OFF 6 In LIVE Mode The PT6 will display...

Страница 2: ...mV displays millivolts and Temperature ESC 6 Release the push button when desired parameter type is displayed SAVEd will display then the PT6 will power OFF II TEMPERATURE UNIT SELECTION The PT6 allow...

Страница 3: ...re in Nitrate NO3 To select the optimal Standard Solutions and to verify calibration see the CONVERSION TABLE below CONVERSION TABLE Standard Solution Value in NO3 N Value in NO3 N1SOL 1 ppm 4 4 ppm N...

Страница 4: ...cts an error in the gain slope it will display Error If this happens clean and condition the sensor then verify you are using approved and properly prepared calibration solutions and recalibrate NOTE...

Страница 5: ...s hold the sensor steady during measurement In HOLD mode real time readings are displayed until the LED is ON solid In LIVE mode real time readings are displayed until the PT6 turns OFF 90 sec in HOLD...

Страница 6: ...y using concentrated standard solutions to increase the nitrate concentration in the sample by a known value optimally double the measured value and then compare the value with the increase in Pen rea...

Страница 7: ...LLING THE NITRATE SENSOR Follow the instructions that come with your replacement sensor Only remove replace the Nitrate sensor in a CLEAN and DRY environment To remove the Nitrate sensor 1 Remove the...

Страница 8: ...ngs ca gov 1 Accuracy of measuring calibration solutions after Temperature Controlled 2 Point Calibration with one order of concentration difference 2 For extended life of the sensor store the PT6 in...

Отзывы: