background image

 

3

 

 

 

 

If LIVE mode is selected you will be prompted to set the auto-off time: The display will alternate between “PUSH CONT” 
and “[value] SEC”. Press and release repeatedly or push and hold to adjust the time out period. The display will scroll from 
140 to 300 seconds, then repeat. When you have selected the desired timeout period, let the PT6 time out. 

SENSOR CONDITIONING 

If the sensor is new and/or has been stored dry, the sensor needs to be conditioned before calibration or measurement. 
1.

 

Remove the protective cap. Rinse the sensor with clean water (preferably DI, RO, or distilled) and pat it dry with a clean soft 
lint-free cloth. 

2.

 

Fill  the  protective  cap  half-full  with  PREPARED  calibration  solution  (with  Nitrate  ISA  added,  See  PREPARING 
CALIBRATION  SOLUTION,  below),  carefully  put  the  cap  back  on  the  sensor  for  conditioning  (using  caution  as  excess 
solution may squirt out). It may take up to an hour for sensor to restore stable readings after dry storage.   

NOTE:  For  best  results,  you  should  always  condition  your  sensor  using  a  PREPARED  calibration  solution  with  a  value 
close to the lower end of the concentration values you will be measuring. 

CALIBRATION 

The Myron L

®

 Company recommends calibrating the PT6 before each series of tests, depending on usage. The PT6 offers 1-point 

or  2-Point  calibration  and  is  calibrated  in  Nitrate  as  Nitrogen  (NO

3

-

-N)  using  1,  10,  100,  or  1000  ppm  Standard  Solutions.  For 

greatest  accuracy  complete  a  2-point  calibration,  selecting  the  closest  Standard  Solutions  above  and  below  your  expected  test 
solution  range  (example:  If  you  are  measuring  test  solutions  in  the  range  of  200  to  400  ppm  NO

3

-

-N,  then  you  would  want  to 

calibrate using 100 ppm and 1000 ppm Standard Solutions). Calibration solutions must be prepared prior to use. 
Calibration  is  always  done  in  Nitrate  as  Nitrogen  (NO

3

-

-N)  even  if  your  PT6  is  set  to  measure  in  Nitrate  (NO

3

-

).  To  select  the 

optimal Standard Solutions and to verify calibration, see the CONVERSION TABLE below. 

 

CONVERSION TABLE 

Standard Solution 

Value in NO

3

-

-N 

Value in NO

3

-

 

N1SOL 

1 ppm 

4.4 ppm 

N10SOL 

10 ppm 

44 ppm 

N100SOL 

100 ppm 

440 ppm 

N1000SOL 

1000 ppm 

4400 ppm 

NOTES 

 

Always start with the lower calibration point first. 

 

For greatest accuracy, keep the temperature of prepared calibration solution(s) the same as the test solution. The 
temperature difference should be within ±2

o

C.  

 

Nitrate sensor should be conditioned prior to performing a calibration (see Sensor Conditioning, above). 

I. PREPARING CALIBRATION SOLUTION 

Calibration solution(s) must be prepared prior to calibration. For each standard solution selected: 
1.

 

Fill a Graduated Cylinder to the 10 mL line with the Standard Solution. 

2.

 

Add 200 

µ

L of the Nitrate ISA solution to the Graduated Cylinder (using the supplied 100 

µ

L pipette). 

NOTE

:  When  using  the  pipette  to  draw,  depress  pipette  to  1st  stop,  then  place  tip  in  ISA  solution  and  release  slowly.  To 

dispense, depress pipette completely to the 2

nd

 stop.  

NOTE

:  Do  not  contaminate  the  pipette  tip  by  touching  it  with  your  fingers  or  setting  it  in  a  dirty  toolbox.  Do  not  dip  the 

pipette tip in the calibration solution as this will cause cross-contamination between solutions. 

3.

 

Replace the cap on the Graduated Cylinder and shake well to mix the prepared calibration solution.  

II. 2-POINT CALIBRATION

  

To perform a 2-Point calibration you will need two prepared calibration solutions, clean water (preferably DI, RO, or distilled), 
and a clean, dry, lint-free cloth. 
1.

 

Remove the protective cap. Rinse the sensor for about 5 seconds with clean water (preferably DI, RO, or distilled) and pat it 
dry with a clean soft lint-free cloth. 

2.

 

Press and release the push button to turn the PT6 ON. 

3.

 

Push and hold the push button to enter menu mode. 

4.

 

Release the button while “CAL” is displayed. The display will alternate between “PUSHnHLD” and “CAL”. 

5.

 

Push and hold the push button. The display will alternate between “1ppm”, “10ppm”, “100ppm”, “1000ppm”, and “ESC”. 

Содержание Ultrapen PT6

Страница 1: ...he sensor submerged 5 In HOLD mode Real time readings are displayed until the LED turns solid ON The display holds the final reading briefly then the PT6 turns OFF 6 In LIVE Mode The PT6 will display...

Страница 2: ...mV displays millivolts and Temperature ESC 6 Release the push button when desired parameter type is displayed SAVEd will display then the PT6 will power OFF II TEMPERATURE UNIT SELECTION The PT6 allow...

Страница 3: ...re in Nitrate NO3 To select the optimal Standard Solutions and to verify calibration see the CONVERSION TABLE below CONVERSION TABLE Standard Solution Value in NO3 N Value in NO3 N1SOL 1 ppm 4 4 ppm N...

Страница 4: ...cts an error in the gain slope it will display Error If this happens clean and condition the sensor then verify you are using approved and properly prepared calibration solutions and recalibrate NOTE...

Страница 5: ...s hold the sensor steady during measurement In HOLD mode real time readings are displayed until the LED is ON solid In LIVE mode real time readings are displayed until the PT6 turns OFF 90 sec in HOLD...

Страница 6: ...y using concentrated standard solutions to increase the nitrate concentration in the sample by a known value optimally double the measured value and then compare the value with the increase in Pen rea...

Страница 7: ...LLING THE NITRATE SENSOR Follow the instructions that come with your replacement sensor Only remove replace the Nitrate sensor in a CLEAN and DRY environment To remove the Nitrate sensor 1 Remove the...

Страница 8: ...ngs ca gov 1 Accuracy of measuring calibration solutions after Temperature Controlled 2 Point Calibration with one order of concentration difference 2 For extended life of the sensor store the PT6 in...

Отзывы: