8

 

 

 

2. UNSTABLE READINGS / SLOW RESPONSE 

A soiled membrane or dry reference junction can cause unstable/slow response of sensors. Rinse the sensor with plenty of water 
and recondition the sensor in 100 ppm standard solution for at least one hour.  
Temperature changes, pH variation of solutions, or ambient electrical noise may also contribute to the issue. The source should be 
identified and corrected.  

SPECIFICATIONS 

 

Nitrate as Nitrogen NO

3

-

-N 

Nitrate 

NO

3

-

 

Temperature 

Ranges 

1-1400 ppm 

4.4-6200 ppm 

0-71˚C / 32-160˚F 

Accuracy

1

 

±10% of the reading

 

±10% of the reading 

± 0.1°C 

Resolution 

0.1 ppm 

0.1 ppm 

0.1°C/F 

pH Range: 3 to 11 pH 
Time to Reading Stabilization: 10 - 180 seconds 
Power Consumption: Active Mode: 33 - 140 mA 
Physical Dimensions: 17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D 
Weight: 54 g or 1.94 oz. 
Case Material: Anodized Aircraft Aluminum with Protective Coating 
Battery Type: One N type, Alkaline 1.5 V 
Calibration Standard Solutions: 1, 10, 100, or 1000 ppm NO

3

-

-N  

Operating / Storage Temperature: 0-40ºC / 32-104ºF 

2

 

Enclosure Ratings: IP67 and NEMA 6 
EN61236-1: 2006 - Annex A: 2008: Electrostatic discharge to the PT6 may cause it to spontaneously power ON.  
In this case, the PT6 will power OFF after several seconds. 

        WARNING!

 These products can expose you to chemicals including Di(2-ethylhexyl)phthalate (DEHP), which is known to the  

        State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov 

1

 Accuracy of measuring calibration solutions after Temperature Controlled 2-Point Calibration with one order of concentration difference. 

2

 For extended life of the sensor, store the PT6 in a cool location (5-25˚C). DO NOT FREEZE! 

EPA Approved Standard Method 4500-NO

3

-

 D* 

This method using the nitrate ion selective electrode is one of EPA approved methods for measuring nitrate in drinking water and 
wastewater, listed in USA Code of Federal Regulations Title 40 Part 136 and Part 141. The PT6 is compatible to follow this 
method. Consult with your regulator for reporting purposes. 
* Standard Methods for the Examination of Water and Wastewater 20

th

 Edition, published by American Public Health Association, 

Washington, D.C., 20005. USA. www.standardmethods.org. 

ACCESSORIES 

STANDARD SOLUTIONS FOR CALIBRATION: 

The ULTRAPEN™ PT6 can be calibrated with 1, 10, 100, or 1000 ppm NO

3

-

-N Standard Solution. 

Order Models: 

 

N1SOL

 

1 ppm NO

3

-

-N (Nitrate as Nitrogen) Standard Solution 

 

N10SOL

 

10 ppm NO

3

-

-N (Nitrate as Nitrogen) Standard Solution 

 

N100SOL

 

100 ppm NO

3

-

-N (Nitrate as Nitrogen) Standard Solution 

 

N1000SOL

 

1000 ppm NO

3

-

-N (Nitrate as Nitrogen) Standard Solution 

IONIC STRENGTH ADJUSTER: 

Order Model: NISASOL 

NITRATE INTERFERENCE SUPPRESSION BUFFER: 

Order Model: NISBSOL 

REPLACEMENT NITRATE SENSOR: 

Order Model: RPT6 

WARRANTY 

The PT6, excluding the sensor and battery, has a one (1) year limited warranty. 
The Nitrate sensor has a six (6) month limited warranty. 
 

MYRON L

®

 COMPANY

 

2450 Impala Drive 

u

 

Carlsbad, CA 92010-7226 

u

 

Phone: +1-760-438-2021 

E-Mail: Customer Service – [email protected] 

u

 

Technical Support – [email protected] 

Website: www.myronl.com 

PT6OM 09/18

 

Содержание Ultrapen PT6

Страница 1: ...he sensor submerged 5 In HOLD mode Real time readings are displayed until the LED turns solid ON The display holds the final reading briefly then the PT6 turns OFF 6 In LIVE Mode The PT6 will display...

Страница 2: ...mV displays millivolts and Temperature ESC 6 Release the push button when desired parameter type is displayed SAVEd will display then the PT6 will power OFF II TEMPERATURE UNIT SELECTION The PT6 allow...

Страница 3: ...re in Nitrate NO3 To select the optimal Standard Solutions and to verify calibration see the CONVERSION TABLE below CONVERSION TABLE Standard Solution Value in NO3 N Value in NO3 N1SOL 1 ppm 4 4 ppm N...

Страница 4: ...cts an error in the gain slope it will display Error If this happens clean and condition the sensor then verify you are using approved and properly prepared calibration solutions and recalibrate NOTE...

Страница 5: ...s hold the sensor steady during measurement In HOLD mode real time readings are displayed until the LED is ON solid In LIVE mode real time readings are displayed until the PT6 turns OFF 90 sec in HOLD...

Страница 6: ...y using concentrated standard solutions to increase the nitrate concentration in the sample by a known value optimally double the measured value and then compare the value with the increase in Pen rea...

Страница 7: ...LLING THE NITRATE SENSOR Follow the instructions that come with your replacement sensor Only remove replace the Nitrate sensor in a CLEAN and DRY environment To remove the Nitrate sensor 1 Remove the...

Страница 8: ...ngs ca gov 1 Accuracy of measuring calibration solutions after Temperature Controlled 2 Point Calibration with one order of concentration difference 2 For extended life of the sensor store the PT6 in...

Отзывы: