2. UNSTABLE READINGS / SLOW RESPONSE
A soiled membrane or dry reference junction can cause unstable/slow response of sensors. Rinse the sensor with plenty of water
and recondition the sensor in 100 ppm standard solution for at least one hour.
Temperature changes, pH variation of solutions, or ambient electrical noise may also contribute to the issue. The source should be
identified and corrected.
SPECIFICATIONS
Nitrate as Nitrogen NO
3
-
-N
Nitrate
NO
3
-
Temperature
Ranges
1-1400 ppm
4.4-6200 ppm
0-71˚C / 32-160˚F
Accuracy
1
±10% of the reading
±10% of the reading
± 0.1°C
Resolution
0.1 ppm
0.1 ppm
0.1°C/F
pH Range: 3 to 11 pH
Time to Reading Stabilization: 10 - 180 seconds
Power Consumption: Active Mode: 33 - 140 mA
Physical Dimensions: 17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D
Weight: 54 g or 1.94 oz.
Case Material: Anodized Aircraft Aluminum with Protective Coating
Battery Type: One N type, Alkaline 1.5 V
Calibration Standard Solutions: 1, 10, 100, or 1000 ppm NO
3
-
-N
Operating / Storage Temperature: 0-40ºC / 32-104ºF
2
Enclosure Ratings: IP67 and NEMA 6
EN61236-1: 2006 - Annex A: 2008: Electrostatic discharge to the PT6 may cause it to spontaneously power ON.
In this case, the PT6 will power OFF after several seconds.
WARNING!
These products can expose you to chemicals including Di(2-ethylhexyl)phthalate (DEHP), which is known to the
State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
1
Accuracy of measuring calibration solutions after Temperature Controlled 2-Point Calibration with one order of concentration difference.
2
For extended life of the sensor, store the PT6 in a cool location (5-25˚C). DO NOT FREEZE!
EPA Approved Standard Method 4500-NO
3
-
D*
This method using the nitrate ion selective electrode is one of EPA approved methods for measuring nitrate in drinking water and
wastewater, listed in USA Code of Federal Regulations Title 40 Part 136 and Part 141. The PT6 is compatible to follow this
method. Consult with your regulator for reporting purposes.
* Standard Methods for the Examination of Water and Wastewater 20
th
Edition, published by American Public Health Association,
Washington, D.C., 20005. USA. www.standardmethods.org.
ACCESSORIES
STANDARD SOLUTIONS FOR CALIBRATION:
The ULTRAPEN™ PT6 can be calibrated with 1, 10, 100, or 1000 ppm NO
3
-
-N Standard Solution.
Order Models:
•
N1SOL
-
1 ppm NO
3
-
-N (Nitrate as Nitrogen) Standard Solution
•
N10SOL
-
10 ppm NO
3
-
-N (Nitrate as Nitrogen) Standard Solution
•
N100SOL
-
100 ppm NO
3
-
-N (Nitrate as Nitrogen) Standard Solution
•
N1000SOL
-
1000 ppm NO
3
-
-N (Nitrate as Nitrogen) Standard Solution
IONIC STRENGTH ADJUSTER:
Order Model: NISASOL
NITRATE INTERFERENCE SUPPRESSION BUFFER:
Order Model: NISBSOL
REPLACEMENT NITRATE SENSOR:
Order Model: RPT6
WARRANTY
The PT6, excluding the sensor and battery, has a one (1) year limited warranty.
The Nitrate sensor has a six (6) month limited warranty.