MENU OPTIONS 

The  PT6  Menu  allows  the  user  to  navigate  through  the  available  function/setting  options.  See  detailed  instructions  below  for 
complete instructions on each function/setting option.  
To get to the menu, simply press and release the push button to turn the PT6 ON, then push and hold the push button. The PT6 
display will alternate between the following options: 

 

CAL – Calibration Menu 

 

FAC CAL – Factory Calibration Menu 

 

ºCºF TEMP – Temperature Unit Selection Menu 

 

ModE SEL – Mode Selection Menu 

 

PAr SEL – Nitrate Parameter Selection Menu 

 

SEnSor 

 – Sensor Check Menu 

 

ESC – Escapes/Exits the Menu and immediately powers the PT6 OFF 

Release the push button while the desired option is displayed. See detailed instructions below. 

ULTRAPEN SETTINGS 

NOTE: 

In all menus, selecting “ESC” immediately powers the PT6 OFF without saving changes to the settings. 

I. NITRATE PARAMETER SELECTION  

The PT6 allows you to select how you wish to display the nitrate concentration readings: Nitrate as Nitrogen (NO

3

-

-N) displayed 

as ppm N, Nitrate (NO

3

-

) displayed as ppm NO3, or millivolt displayed as mV.   

To select a parameter for display:  
1.  Press and release the push button to turn the PT6 ON.  
2.  Push and hold the push button to enter menu mode.  
3.  Release the push button while “PAr SEL” is displayed.   
4.  The display will alternate between “PUSHnHLD” and “PAr SEL”.  
5.  Push and hold the push button. The display will alternate between the following: 

 

“N” (displays Nitrate as Nitrogen and Temperature) 

 

“NO3” (displays Nitrate and Temperature) 

 

“mV” (displays millivolts and Temperature) 

 

“ESC”  

6.  Release the push button when desired parameter type is displayed. 

 

“SAVEd” will display, then the PT6 will power OFF. 

II. TEMPERATURE UNIT SELECTION 

 

The PT6 allows you to select the type of units used for displaying temperature; ˚C (Celsius) or ˚F (Fahrenheit).  
To select the Temperature Units: 
1.

 

Press and release the push button to turn the PT6 ON.  

2.

 

Push and hold the push button to enter menu mode.  

3.

 

Release the push button while “˚C˚F TEMP” is displayed.   

4.

 

The display will alternate between “PUSHnHLD” and “˚C˚F TEMP”. 

 

Push and hold the push button. The display will alternate between “˚C” (Celsius), “˚F” (Fahrenheit), and “ESC”. 

5.

 

Release the push button when desired unit is displayed. 

6.

 

“SAVEd ˚C” or “SAVEd ˚F” will display, then the PT6 will power OFF. 

III. MODE SELECTION 

The PT6 allows you to select between two measurement modes; Live mode or Hold mode.  

HOLD 

mode  (Default) 

– 

Real-time  readings  are  displayed  until  the  LED  turns  solid  ON.  The  display  holds  the  final  reading 

briefly then the PT6 turns OFF. 

LIVE 

mode 

– 

Real-time readings are displayed continuously for up to 5 minutes. A press and release of the push button will turn 

your PT6 OFF immediately. User may select auto-off timeout period from 140 to 300 seconds. 
To select the mode: 
1.  Press and release the push button to turn the PT6 ON. 
2.  Push and hold the push button to enter menu mode.  
3.  Release the push button while “ModE SEL” is displayed.   
4.  The display will alternate between “ModE SEL” and “PUSHnHLD”.  
5.  Push and hold the push button. The display will alternate between “Hold”, “LIVE” and “ESC”. 
6.  Release the push button when the desired mode is displayed.  

 

If Hold mode is selected: “SAVEd” will display, then the PT6 will power OFF. 

Содержание Ultrapen PT6

Страница 1: ...he sensor submerged 5 In HOLD mode Real time readings are displayed until the LED turns solid ON The display holds the final reading briefly then the PT6 turns OFF 6 In LIVE Mode The PT6 will display...

Страница 2: ...mV displays millivolts and Temperature ESC 6 Release the push button when desired parameter type is displayed SAVEd will display then the PT6 will power OFF II TEMPERATURE UNIT SELECTION The PT6 allow...

Страница 3: ...re in Nitrate NO3 To select the optimal Standard Solutions and to verify calibration see the CONVERSION TABLE below CONVERSION TABLE Standard Solution Value in NO3 N Value in NO3 N1SOL 1 ppm 4 4 ppm N...

Страница 4: ...cts an error in the gain slope it will display Error If this happens clean and condition the sensor then verify you are using approved and properly prepared calibration solutions and recalibrate NOTE...

Страница 5: ...s hold the sensor steady during measurement In HOLD mode real time readings are displayed until the LED is ON solid In LIVE mode real time readings are displayed until the PT6 turns OFF 90 sec in HOLD...

Страница 6: ...y using concentrated standard solutions to increase the nitrate concentration in the sample by a known value optimally double the measured value and then compare the value with the increase in Pen rea...

Страница 7: ...LLING THE NITRATE SENSOR Follow the instructions that come with your replacement sensor Only remove replace the Nitrate sensor in a CLEAN and DRY environment To remove the Nitrate sensor 1 Remove the...

Страница 8: ...ngs ca gov 1 Accuracy of measuring calibration solutions after Temperature Controlled 2 Point Calibration with one order of concentration difference 2 For extended life of the sensor store the PT6 in...

Отзывы: