background image

 

5

 

 

 

NITRATE MEASUREMENT  

I. MEASUREMENT PREPARATION 

The following steps should be taken prior to performing a measurement with your PT6: 

1.

 

Perform  a

 

calibration  check  with  calibration  solution,  and  if  the  reading  is  off  more  than  10%,  calibrate  the  PT6.  (See 

CALIBRATION). 

2.

 

Prepare test solution:  
a.

 

Fill the Graduated Cylinder to the 10 mL line with solution to be tested. 

b.

 

Add 200 

µ

L of the Nitrate ISA Solution to the Graduated Cylinder (using the supplied 100 

µ

L pipette). 

NOTE

:  When  using  the  pipette  to  draw,  depress  pipette  to  1

st

  stop,  then  place  tip  in  solution  and  slowly  release.  To 

dispense, depress pipette completely to the 2

nd

 stop.  

NOTE

: Do not contaminate the pipette tip by touching it with your fingers or setting it in a dirty toolbox. Do not dip the 

pipette tip in the calibration solution as this will cause cross-contamination between solutions. 

c.

 

Replace the cap on the Graduated Cylinder and shake well to mix the prepared test solution.  

II. NITRATE MEASUREMENT 

For  highest  accuracy,  the  temperature  of  the  calibration  solution(s)  should  be  the  same  as  the  temperature  of  the  test 
solution(s). The temperature difference should be no more than ±2

o

C/±3.6

o

F. 

1.

 

Remove the protective cap. Rinse the sensor for about 5 seconds with clean water (preferably DI, RO, or distilled) and pat it 
dry with a clean soft lint-free cloth. 

2.

 

Press and release the push button to turn the PT6 ON. 

3.

 

Dip the PT6 sensor in the prepared test solution and swirl it around to remove any air bubbles. 

4.

 

Hold the PT6 sensor steady in the solution while the measurement is taking place. 

The following table explains what the LED Indicator Light signals indicate and gives the duration of each signal: 

LED SIGNAL

 

ACTION

 

DURATION

 

Rapid Flashing 

Dip sensor in solution and swirl to remove bubbles. 

6 sec 

Slow Flashing 

Measurement in process; hold the sensor steady during measurement. 
In 

HOLD

 mode real-time readings are displayed until the LED is ON solid. 

In 

LIVE

 mode real-time readings are displayed until the PT6 turns OFF. 

90 sec in 

HOLD

 mode 

Up to 300 sec in 

LIVE

 mode 

Solid Light (

HOLD

 mode only) 

Measurement is complete. Values are displayed until the PT6 turns OFF. 

6 sec 

5.

 

Remove the sensor from the solution and rinse with clean water (preferably DI, RO, or distilled) and pat it dry with a clean 
soft lint-free cloth. Use caution not to touch the ISE membrane.  

6.

 

Fill the protective cap half full with prepared calibration solution. Carefully put the cap back on the sensor for conditioning 
(using caution as excess solution may squirt out). 

NOTE

: You can turn OFF the unit at any time during LIVE measurement with a simple press and release of the push button. 

 

For  best  results  when  measuring  solution  at  the  extremes  of  the  specified  temperature  or  nitrate  range  allow  the  PT6  to 

equilibrate by submerging the sensor in the sample solution for 1 minute prior to taking a measurement. 

 

 

ALWAYS

 make sure the nitrate sensor is completely submerged in the solution being tested. 

 

For multiple samples, regularly perform a calibration check with calibration solution between sample measurements. 

 

III. INTERFERENCE

 

Other anions in solution may interfere with nitrate measurement, depending on the species and their relative levels to nitrate. The 
impact of interference ions on measurements can be expressed by the Nikolsky equation. 

𝐸

=

𝐶𝑜𝑛𝑠𝑡𝑎𝑛𝑡

+

!

.

!"!

!"

!

!

!

log

 

(

𝑎

 

+

𝐾

!

 

𝑎

!

!

!

!

!

Where    

𝐸

- the voltage sensor measures 

 

𝑎

 

- the activity of nitrate 

 

𝑎

!

- the activity of interference ion 

 

𝑧

!

- the charge of the interference ion with sign 

 

𝐾

!

 

–the Selectivity coefficient of the interference ion for nitrate measurement  

 
 

 

 

Generally, interference ions may cause positive errors for the measurement. The higher selectivity coefficient the interference ion 
has, the higher error it may cause at a given concentration. The approximate selectivity coefficients of PT6 for some anions are 

 

SO

4

2-

 

Cl

-

 

CH

3

COO

-

 

K(NO

3

-

/x) 

0.00016  0.007  0.0007 

Содержание Ultrapen PT6

Страница 1: ...he sensor submerged 5 In HOLD mode Real time readings are displayed until the LED turns solid ON The display holds the final reading briefly then the PT6 turns OFF 6 In LIVE Mode The PT6 will display...

Страница 2: ...mV displays millivolts and Temperature ESC 6 Release the push button when desired parameter type is displayed SAVEd will display then the PT6 will power OFF II TEMPERATURE UNIT SELECTION The PT6 allow...

Страница 3: ...re in Nitrate NO3 To select the optimal Standard Solutions and to verify calibration see the CONVERSION TABLE below CONVERSION TABLE Standard Solution Value in NO3 N Value in NO3 N1SOL 1 ppm 4 4 ppm N...

Страница 4: ...cts an error in the gain slope it will display Error If this happens clean and condition the sensor then verify you are using approved and properly prepared calibration solutions and recalibrate NOTE...

Страница 5: ...s hold the sensor steady during measurement In HOLD mode real time readings are displayed until the LED is ON solid In LIVE mode real time readings are displayed until the PT6 turns OFF 90 sec in HOLD...

Страница 6: ...y using concentrated standard solutions to increase the nitrate concentration in the sample by a known value optimally double the measured value and then compare the value with the increase in Pen rea...

Страница 7: ...LLING THE NITRATE SENSOR Follow the instructions that come with your replacement sensor Only remove replace the Nitrate sensor in a CLEAN and DRY environment To remove the Nitrate sensor 1 Remove the...

Страница 8: ...ngs ca gov 1 Accuracy of measuring calibration solutions after Temperature Controlled 2 Point Calibration with one order of concentration difference 2 For extended life of the sensor store the PT6 in...

Отзывы: