Table 4
Resistance to Electromagnetic Interference
IMMUNITY test
Standard-
Specification
IEC 60601-1-2:2014
Test level
IEC 60601-1-2:2014
Compliance level for OW100
®
and electromagnetic
environment
– guidance
Electrostatic
discharge
IEC 61000-4-
2:2008
Contact discharge: ±8kV
and air discharge: ±2kV,
±4kV, ±8kV, ±15kV
compliance *
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least 30
%.
High frequency
electromagnetic
fields
IEC 61000-4-
3:2010
80MHz to 2.7GHz @
10V/m
80% AM at 1kHz
compliance *
High frequency
electromagnetic
fields in direct
neighborhood of
wireless
communication
devices
IEC 61000-4-
3:2010
See table 2
compliance *
Electrical fast
transient/bursts
IEC 61000-4-
4:2012
±2kV for mains cables &
±1kV for I/O cables @
100 kHz repetition
frequency
compliance *
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surges
IEC 61000-4-
5:2014
±0.5kV, ±1kV & ±2kV (L-
PE and N-PE) and
±0.5kV & ±1kV (L-N) @
0, 90 and
270˚
compliance *
Mains power quality should
be that of a typical
commercial or hospital
environment.
Disturbances on
cables, induced by
high frequency fields
IEC 61000-4-
6:2013
150kHz to 80MHz @
3Vrms; 6V in ISM bands
between 150kHz-80MHz
80% AM at 1kHz
compliance *
Power frequency
(50/60 Hz)
Magnetic fields
IEC 61000-4-
8:2009
30A/m @ 50Hz or 60Hz
compliance *
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
*no change of an essential performance criteria