MTS_OW100_IFU-orthogold100-US-K182682_A
Energy Setting Values:
Per IEC 63045:2017 Draft: This International Standard specifies measurable parameters
which can be used in the declaration of the acoustic output of extracorporeal unfocused or
weakly focused pressure pulse sources.
OW100
®
Generator setting
(Energy)
p
c
[Mpa]
p
r
[Mpa]
P
II
T
[mJ/mm
2
]
P
II
P
[mJ/mm
2
]
1
0.43
-0.17
0.00017
0.00015
2
0.78
-0.24
0.00071
0.00056
3
1.21
-0.32
0.00148
0.00117
4
1.69
-0.40
0.00254
0.00205
5
2.23
-0.49
0.00395
0.00328
6
2.81
-0.57
0.00580
0.00493
7
3.44
-0.66
0.00814
0.00708
8
4.10
-0.75
0.01100
0.00977
9
4.79
-0.84
0.01439
0.01299
10
5.49
-0.93
0.01829
0.01670
11
6.20
-1.02
0.02260
0.02078
12
6.89
-1.12
0.02719
0.02504
13
7.55
-1.22
0.03186
0.02925
14
8.17
-1.32
0.03640
0.03311
15
8.73
-1.42
0.04054
0.03635
16
9.27
-1.52
0.04403
0.03870
Figure
33
Dependence of the pressure pulse parameters on the generator