35
MTS_OW100_IFU-orthogold100-US-K182682_A
CAUTION
AM-, FM, TV- antennas
Observe the essential requirements of the OW100
®
if the device
is used closer than 1.5km (0.9 miles) from an AM-, FM- or a TV-
antenna.
If the OW100
®
1. Shows wrong energy parameters
2. Releases unintended shockwaves,
resulted by electromagnetical interferences,
STOP the treatment directly and unplug the device from the
power socket.
WARNING
Accessory!
The use of other accessories, transformers, and other leads than
those specified and provided by the manufacturer of this device
may result in increased electromagnetic emissions or reduced
electromagnetic immunity, which can result in an incorrect
operation.
Use only the accessories and cables listed in Table 1
CAUTION
Electromagnetic Interactions!
If the device is used in combination with other electrical
equipment the entire configuration must be in accordance with EN
60601-1-1. Consult a specialist in case of doubt.
The OW100
®
may not be used in the vicinity of, or together with
other equipment in a stacked arrangement.
If it is unavoidable to stack the device in close proximity to other
equipment, the OW100
®
should be carefully observed, in order
detect any adverse effects among the devices installed in the
arrangement.
NOTE
The OW100
®
is intended for use in an electromagnetic
environment, in which HF interferences are controlled. You can
help to reduce electromagnetic interference by obtaining the
minimum safe distance between mobile HF telecommunication
devices (transmitters) and the OW100
®
– dependent on the
output power of the communication device specified in table 2
CAUTION
Electromagnetic Interactions!
If the device is used in combination with other electrical
equipment the entire configuration must be in accordance with EN
60601-1-1. Consult a specialist in case of doubt.
The OW100
®
may not be used in the vicinity of, or together with
other equipment in a stacked arrangement.
If it is unavoidable to stack the device in close proximity to other
equipment, the OW100
®
should be carefully observed, in order
detect any adverse effects among the devices installed in the
arrangement.