WARNING
Electrical hazard danger
Make sure that the designated wall socket has a PE connection!
CAUTION
Standards of the local power supply
Ensure that the power connector of the
OW100
®
fits to the existent power socket.
3.1
Applicator
OW100
®
applicator should be exchanged after its use life
(when display shows “0”). At first
delivery, it is included as ordered.
Connect the applicator
The applicator is delivered in a special packaging. Store package at a safe place for
optionally shipping after usage.
CAUTION
Applicator hose
Do not kink the connecting hose of the applicator. Kinking
the hose repeatedly may result in damage to the internal
components that control applicator functions. If hose
appears to have been kinked and applicator functions fail,
replace the applicator immediately.
WARNING
Inspect the water membrane for tears or leaks
Check water membrane of the applicator for damage before setting
into operation. If the membrane is damaged, the applicator should not
be used and must be replaced immediately.
To connect the applicators to the OW100
®
proceed as follows:
1. Switch on the OW100
®
into ready mode (If the device is in stand-by-mode (screen dark,
power switch illuminated and the lower LED in touch-wheel illuminated), press confirmation
button in the touch wheel for 2 seconds). The OW100
®
starts the self test and ends after
successful testing in the operation mode.
2. Make sure the position of the plug to the socket at the device is correct.