Running the Example HCF Test
80
Landmark™ Test System Operation
2. Verify initial force tuning parameters.
A. To open the Tuning: Ch1 Force window, select
Station Manager >Display >
Station Setup > Channels > Landmark 1 (or other depending on your
configuration) > Force
and then click the Tuning Fork button.
B. Review the following parameters for this hard specimen test. The following is a good
place to start for this Example test with a hard specimen.
l
P Gain
— 1
l
I Gain
— 1
l
D Gain
— 0
l
F Gain
— 0
l
FL Filter
— Set to maximum positive value.
3. Disable manual command.
This allows the function generator to control the system.
A. In the Station Manager main window, click the Manual Command button. The Manual
Commands window appears.
B. Click to clear (uncheck) the
Enable Manual Command
check box.