Running the Example HCF Test
66
Landmark™ Test System Operation
3. Set the Force Lower (Compression) Limit.
For specimen installation, select a low compression limit to reduce the chance of specimen
damage or injury. While -50 N is shown here as an example, you should use a limit that
adheres to your lab's specimen installation guidelines.
A. Enter
-50 N
in the
Lower Limit
text box (a negative number indicates compression).
Be sure to press
Enter
after typing in the limit.
B. Once again, select
Interlock
for the
Action
.
Zero the Force Signal
1. Select
Exclusive Control
.
In the
Station Manager
application main window, select the
Exclusive Control
check box.
This allows control to come from either the computer software or the handset, but not both.