background image

Medical Test Supply 

 

CMS-50FF 

 

   

           

Figure 6.Alarm Setting Menu     

c. Pulse sound indication setting

   

Click  the  button  to  select  “Pulse  Sound”,  and  then  Press  the  button  to  choose  to  have  the  Pulse 

Sound (heart beat) alarm “on” or “off”. 

d. Exit the Alarm settings

 

Click the button to select “EXIT”, then Press the button to exit the Alarm Settings Menu. 

c)

Clock setting 

  

 

On the main menu interface, Click the button to select “Clock”, then enter the clock setting interface 

by press the button. 

 

 

Figure 7. Clock Setting Menu 

a.

 

When entering the clock setting menu, the menu choice bar would be on the item of  “set time”, 

and  the  state  would  always  be  “no”  whenever  it  enters  the  clock  setting  menu  on  the  purpose  of 

avoiding unexpected changes of time due to improper operation.  You can change the state by press 

the button, choose “yes” to reset the time, choose “no” to forbid time resetting. 

b. 

Click the button to select the parameter that you want to change, then adjust the data by press the 

button. 

c.

 

  Click the button to select  “Exit”, then exit the clock setting menu by press the button.  If you 

have reset the time or date, when exiting the clock setting menu, firstly the renewed time and date 

Содержание CMS-50FF

Страница 1: ...ty and performance issues and any monitoring abnormality human injury and equipment damage due to users negligence of the operation instructions The manufacturer s warranty service does not cover such...

Страница 2: ...cations 4 4 1Mainperformance 4 4 2MainParameters 4 5Installation 5 5 1Viewofthefrontpanel 5 5 2Wristbandinstallationandprobeconnection 5 5 3Accessories 6 6OperatingGuide 6 6 1Applicationmethod 6 6 2At...

Страница 3: ...attery plastic bags foams and paper boxes should follow the local laws and regulations Please check the packing before use to make sure the device and accessories are totally in accordance with the pa...

Страница 4: ...d default setting for good This function could be turned on through menu operation if you need Please check the chapter 6 1 as a reference The device may not work for all patients If you are unable to...

Страница 5: ...95 c Atmospheric pressure 500hPa 1060hPa Operating Environment a Temperature 10 40 b Relative Humidity 30 75 c Atmospheric pressure 700hPa 1060hPa 3 Principle Principle of the Oximeter is as follows A...

Страница 6: ...rs A Measurement of SpO2 Measuring range 0 100 Accuracy When the SpO2 measuring range is 70 100 the permission of absolute error is 2 Below 70 unspecified B Measurement of pulse rate Measuring range 3...

Страница 7: ...ication light When the device is in the state of battering charging the indication light appears to be orange and when the battery status is full the light turns to green 5 2 Wristband installation an...

Страница 8: ...by our company never replace it with the similar ones by other manufacturers b Put the finger into the probe c Turn on the device by long pressing the button on the panel d Do not shake the finger and...

Страница 9: ...order to enter the clock interface and the device would return to the measuring interface by Press the button for about 10 seconds again D Pause alarm a Alarm including the alarm of measure data s goi...

Страница 10: ...LM HI SpO2 low limit SpO2 ALM LO Pulse rate high limit PR ALM HI Pulse rate low limit PR ALM LO Press the button and hold to adjust the selected limit to the desired higher value and release the butto...

Страница 11: ...lock Setting Menu a When entering the clock setting menu the menu choice bar would be on the item of set time and the state would always be no whenever it enters the clock setting menu on the purpose...

Страница 12: ...Figure 8 SpO2 program Figure 9 SpO2 Review program If the users choose to turn on the synchronizing display function on computer it would probably take several seconds for the data to appear on the co...

Страница 13: ...d to your computer c In the state of storing it is not applicable for the users to upload the stored date to computer d When the upload of stored data is finished the menu choice bar will move to Exit...

Страница 14: ...digo blue or carbon monoxide hemoglobin COHb or methionine Me Hb or thiosalicylic hemoglobin and some with icterus problem the SpO2 determination by this monitor may be inaccurate C The drugs like dop...

Страница 15: ...The batteries are drained or almost drained 2 The device s malfunction 1 Please recharge the battery 2 Please contact the local service center The display is off suddenly 1 The device s malfunction 2...

Страница 16: ...Function Specification Information Display Mode The Pulse Oxygen Saturation SpO2 2 digit digital OLED display Pulse Rate PR 3 digit digital OLED display Pulse Intensity bar graph bar graph OLED displ...

Страница 17: ...ent Voltage 3 7 rechargeable lithium battery 1 Battery working life Charge and discharge no less than 500 times Power Adapter Input Voltage 100 to 240 VAC 50 60 Hz Output voltage 5 VDC Output current...

Страница 18: ...Medical Test Supply CMS 50FF...

Отзывы: