Medical Test Supply
CMS-50FF
As to the fingers which are too thin or too cold, it would probably affect the normal measure
of the patients' SpO
2
and pulse rate, please clip the thick finger such as thumb and middle finger
deeply enough into the probe.
The pulse oximeter can be used to adult or infant. Whether the device is used to adult or infant,
it depends on the probe selected.
Whether the device is used to adult or infant, it depends on the probe selected.
The update period of data is less than 5 seconds, which is changeable according to different
individual pulse rate.
Please read the measured value when the waveform on screen is equably and steady-going,
this measured value is optimal value. And the waveform at the moment is the standard one.
If some abnormal conditions appear on the screen during test process, pull out the finger and
reinsert to restore normal use.
The device has normal useful life for three years since the first electrified use.
This device has the function of alarming; users can check on this function according to chapter
6.1 as a reference.
The device has the function of limits alarming, when the measured data is beyond the highest
or lowest limit, the device would start alarming automatically on the premise of the alarming
function is on.
The device has the function of alarming, this function can either be paused, or closed (default
setting) for good. This function could be turned on through menu operation if you need. Please check
the chapter 6.1 as a reference.
The device may not work for all patients. If you are unable to achieve stable readings,
discontinue use.
2 Overview
The pulse oxygen saturation is the percentage of HbO
2
in the total Hb in the blood, so-called the O
2
concentration in the blood. It is an important bio-parameter for the respiration. A number of diseases
relating to respiratory system may cause the decrease of SpO
2
in the blood, furthermore, some other
causes such as the malfunction of human body's self-adjustment, damages during surgery, and the
injuries caused by some medical checkup would also lead to the difficulty of oxygen supply in
human body, and the corresponding symptoms would appear as a consequence, such as vertigo,
impotence, vomit etc. Serious symptoms might bring danger to human's life. Therefore, prompt
information of patients' SpO
2
is of great help for the doctor to discover the potential danger, and is of
great importance in the clinical medical field.
The Pulse Oximeter features in small volume, low power consumption, convenient operation and
being portable. It is only necessary for patients to put one of his fingers into a probe for diagnosis, and
a display screen will directly show the measured value of pulse oxygen saturation with the high
veracity and repetition.
2.1 Features
A.
Operation of the product is simple and convenient.
B.
The product is small in volume, light in weight and convenient in carrying.
C.
Low power consumption
Содержание CMS-50FF
Страница 18: ...Medical Test Supply CMS 50FF...