Medical Test Supply
CMS-50FF
Figure 3.
5.3 Accessories
A. a
wristband
B.
a User Manual
C.
a power adapter
D. a
data line
E.
a disk (PC software)
F.
An adult-oximeter probe
(
Model:CMS04
)
An infant-oximeter probe (May purchase selectively)
6 Operating Guide
6.1 Application method
A.
Install the wristband according to the instructions of chapter 4.2.
a)
Put the suitable probe into the jack on the right side of the oximeter. (The probe is limited to be
produced by our company; never replace it with the similar ones by other manufacturers).
b)
Put the finger into the probe.
c)
Turn on the device by long pressing the button on the panel.
d)
Do not shake the finger and keep the patient in a stable state during the process.
e)
The data can be read directly from the screen on the measuring interface.
Fingernails and the luminescent tube should be on the same side.
If the alarm function is on, the device will provide medium-priority alarm signal when
probe or finger is out. Intermittent alarm will occur and the user interface presents "FINGER
OUT".
Medium priority indicating that prompt operator response is required.