14) The display now reads
LoAd3
or
LoAd4
.
Repeat steps 12 and 13
for additional span points
press
TARE
or press
ZERO
if you are finished.
After LoAD4 has passed, the 3460
will automatically advance to the next
step.
/RDG
/RDG
or
TARE
To enter another
span point
15) The display shows
CAL’d
to indicate that the
calibration was successful.
16) Press
TARE.
The display
flashes
C-CAL
followed
by the C-CAL number.
Make a note of this number.
If the X1000 annunciator is lit,
also make a note of that.
This number is used to calibrate when
test weights are unavailable. C-CAL
does not replicate multipoint
calibrations.
17) Press
ZERO
to store the
calibration constants, or
after eight seconds, the
calibration constants will
automatically stored. After
the store message is
displayed, the scale goes
to the next item on the cal
menu,
SEtuP
.
18) Press
ZERO
to exit the Cal
menus and start up the
standard weight display.
The 3460 is ready for use.
19) After ensuring the
calibration is acceptable,
replace the hex seal screw.
If the scale is being used as a Legal-
for-Trade device, place a lead-wire
seal through the cal screw over to
the adjacent screw.
&&D/
Example Value only.
VWRUH
ZERO
ZERO
&$/G
6HWYS
TARE
Flashing
ZERO
LO BATT
CAL
X1000
MSI-3460
Model No.
To skip the C-CAL step
28
MSI-3460 Operator’s Manual
Figure 5-3. Standard Calibration Procedure, Continued.