MAINTENANCE
23
6. MAINTENANCE
DISCONNECT THE CENTRIFUGE FROM THE POWER SUPPLY
User maintenance is minimal, being confined mainly to cleanliness and care of
rotors and accessories. Basic equipment maintenance must be carried out by a
qualified/competent electrician.
6.1 Cleanliness
Ensure that the body of the centrifuge is cleaned periodically to maintain a good
appearance. It is particularly essential to clean out the centrifuge bowl thoroughly,
especially when corrosive materials have been used. A soft cloth dampened in a
solution of mild detergent and warm water will serve for both purposes.
6.2 Drive Shaft
The centrifuge drive shaft should be cleaned periodically with a solvent to remove
excessive grease. When clean, the shaft should be re-lubricated lightly with
petroleum jelly.
6.3 Prevention Of Corrosion
Simple corrosion can usually be detected by eye and appears in the form of
pitting, or white, fluffy deposits on the surface of aluminium rotors. To prevent
corrosion, take care to avoid scratching the surface of rotors and/or accessories.
After use wash rotors, metal buckets, cups, carriers and adaptors in warm soapy
water, dry and store, inverted, in a dry place. Rotors should be protected with MSE
Rotor Spray (Cat. No.17341.151), which is an anti-corrosion and
de-waterising agent.
In addition, rotors should be periodically inspected by a responsible person so that
corrosion can be detected at the onset. Contact MSE (UK) Ltd Technical Support
for further advice on this subject.
6.4 O-Rings
Ensure O-Rings are lightly coated with silicone grease. Check O-Rings regularly
for cuts and abrasions. Seals should be replaced as necessary, or at least once a
year.
6.5 Sterilisation
Except for the 8-place swing-out rotor Cat. No. 43121.111, all rotors, buckets,
sealing caps (O-Rings removed) and the bowl may be sterilised by autoclaving at
120°C. See section 6.9 for bowl removal procedure. If the means available for
disinfection of certain microbiological agents are inadequate, the safety officer
should be consulted and appropriate steps taken
6.6 Disinfecting
The following cleaning fluids may be used
TERMINEX 2 (Available from Arrow Chemicals Ltd.)
VIRKON (Available from Antec International)
These cleaning agents if used as instructed by the manufacturer should not be
harmful to this product, or accessories supplied for use with this product.
WARNING: SOLVENTS OR GRITTY CLEANERS SHOULD NEVER BE USED
EXCESSIVE USE OF WATER SHOULD ALSO BE AVOIDED
6.7 Inspection Or Renewal Of Motor Brushes