![MSE CENTAUR 2 Скачать руководство пользователя страница 22](http://html.mh-extra.com/html/mse/centaur-2/centaur-2_instruction-manual_1846881022.webp)
OPERATION
22
5.6 Lid Safety Circuit
The Lid Safety Circuit has a dual function:
i.
It prevents power being switched to the motor until the lid is properly closed.
ii. It imposes a delay of approximately three and a half minutes in the event of a
tacho failure before the Lid Unlocked indicator comes on and the lid can be
opened.
Two micro-switches in the lid form part of the lid safety circuit. These can be
checked for correct operation as follows:
i.
Press Power switch on and open the lid. The Lid Unlocked indicator lamp
should be off.
ii. Close the lid. The Lid Unlocked indicator lamp should come on. If the lamp
does not come on, check that the lid is firmly closed as this is the most likely
cause of malfunction .
iii. Switch the Timer to Hold or select a Time. The Lid Unlocked indicator should
go off. If this does not occur, then the Lid Locked microswitch has not
operated. Again, the most likely cause is improper closing of the lid.