INSTALLATION
13
3.3 Recommendation For Securing Centrifuge
A Fixing Kit for securing the Centaur 2 centrifuge to a suitably sound worktop is for
securing centrifuge available from MSE as an optional accessory. It consists of an
anchor bracket which is secured to the bench, or table, by two screws with nuts
and washers; and two screws which screw into the body of the centrifuge to
secure it to the bracket. See Figure 3.
Figure 3
Securing Centrifuge to work surface using Bench fixing Kit
(Part No. 41159.188)
If the restraint kit is not used to secure the centrifuge on the bench then, it is the
recommendation of BS EN61010-2-020 that a clearance of 300mm is allowed
around the base of the centrifuge when in the final operating position.
NOTE: Laboratory management procedures should require that no person or any
hazardous materials are within a 300mm boundary while the centrifuge is
operating.
Figure 3a Recommended Clearances
Note: 75mm if solid surface (i.e. wall) otherwise 300mm