GENERAL DESCRIPTION
10
2.2 Centrifuge Bowl
The centrifuge bowl is easily removable for cleaning or disinfection purposes see
section 6 (Maintenance).
2.3 Opening Lid Manually
An electrical interlock prevents the lid being opened while the rotor is turning, or
when the power is disconnected. If access to the rotor is required during a power
failure, insert the Lid Lock Release Key (supplied with accessories) to its fullest
extent, into the small hole on the top of the lid, near the front; move the point
towards the rear against a slight spring pressure to actuate the lid lock
mechanism. Press the lid actuator to the right and open the lid.
The lid is friction-controlled, to remain open at any intermediate position desired.
Front
Figure 2.
Manual Over-ride of Lid Lock
2.4 Specifications
Dimensions
Height
336mm (with lid open - 470mm)
Width
436mm
Depth
487mm
Weight
36kg (without rotor)
Power Supply
Input voltage and frequency.
Cat. No. MSB020.CX1.5
220/240V
50/60 Hz.
Cat. No. MSB020.CX1.1
110/120V
50/60 Hz.
Operating
Humidity
MRH 80% up to 31
°
C decreasing linearly to 50% RH at 40
°
C
Altitude
Up to 2000M
NOTE: Motor, as fitted, contains overtemperature cut-out device.