Appendix
50
8.2
Specifications
8.2.1
SECTOR Imager
NOTE: All performance information including standard signal levels, detection limits, dynamic range, and instrument noise
is based on 150
µ
L read volumes in 96-well MULTI-ARRAY plates using standard plate read parameters.
Unless noted, specifications apply to all models of the SECTOR Imager 2400/2400A and the SECTOR Imager 6000.
8.2.2
Scientific Performance
Table 8:2 Scientific performance
SECTOR Imager 6000
SECTOR Imager 2400/2400A
Standard Signal Levels
Signals generated by Free TAG, 15,000 [Cat # R94AA]
Signals generated by PQ Low Control [Cat # R92TE-2]
12,000–18,000 counts
25–100 counts
12,000–18,000 counts
25–100 counts
Dynamic Range
10
6
10
6
Electronic Noise
Standard Deviation Dark Noise (depends on plate type)
0–16 counts
0–16 counts
8.2.3
Environmental Specifications
Suggested Operational Temperature Range
20–26
°
C (68–78
°
F)
Suggested Operational Humidity Range
10–80% non-condensing
Storage Temperature Range
5–40
°
C (41–104
°
F)
Storage Humidity Range
5–85% non-condensing
Ambient Light
<2000 Lux
8.2.4
Power Requirements
SECTOR Imager 2400/2400A (models 1250/1251)
Voltage
100–240 V~ 50/60 Hz
Current
3.7–2.0 A
SECTOR Imager 6000 (model 1200)
Voltage
100 / 120 / 220 / 240 V~ 50/60 Hz
Current
3.3 / 2.7 / 1.5 / 1.4 A
SECTOR Imager 6000 (model 1201)
Voltage
100–240 V~ 50/60 Hz
Current
3.5A–2.8A @ 100–240 V~
Class I, Installation (Overvoltage) Category II Device
Содержание SECTOR Image 2400
Страница 1: ...SECTOR Imager Models2400 6000 INSTRUMENT MANUAL www mesoscale com ...
Страница 6: ...6 1 Introduction ...
Страница 9: ...9 2 ImportantInformation ...
Страница 15: ...15 3 SystemDescription ...
Страница 28: ...28 4 Installation ...
Страница 30: ...30 5 QuickStart ...
Страница 36: ...Quick Start 36 Figure 5 8 SECTOR Imager window Run Options dialog box ...
Страница 38: ...38 6 UsingtheSECTORImager ...
Страница 42: ...42 7 Maintenance ...
Страница 45: ...45 8 Appendix ...
Страница 54: ...54 9 TechnicalSupport ...
Страница 56: ...Technical Support 9 2 ...