3.2 Mounting
Allow convenient access to the monitor in order to input local configurations and calibrate
via the magnetic switch on the cover.
Leave sufficient room to attach cabling. Keep cable run distances below specified limits.
Although the Model IR4000S is RFI-resistant, where possible mount it away from heavy
power cabling, radio transmitters, or similar equipment.
Locate the IR4000S away from concentrated sources of heat or light. It should be
mounted in areas that are free from excessive vibration and as free as possible from
wind, dust, water, and direct sunlight.
YEAR OF CONSTRUCTION
MODEL IR4000M
P/N 32580
CONFIG
MAN
S/N
0518
0539
CALIBRATION PROCEDURE
1. HAVE DETECTOR IN CLEAN AIR
2. APPLY MAGNET, AND WAIT FOR "AC"
ON DISPLAY
3. REMOVE MAGNET, WHEN d# IS
DISPLAYED, APPLY MAGNET TO
SELECT DETECTOR #
4. REMOVE MAGNET, FLASHING "AC"
IS DISPLAYED
5. WHEN "AC" IS STEADY, APPLY GAS
6. WHEN GAS IS DETECTED, "CP" IS
DISPLAYED
7. WHEN CALIBRATION IS COMPLETE,
"CC" IS DISPLAYED
8. REMOVE GAS
ALARM LED
WARN LED
Figure 4. IR4000S with IR400 Mounting Dimensions, in inches [mm]
NOTE:
Nameplate is for reference purposes only and may not contain up-to-date
information.