MRU Instruments, Inc.
USER MANUAL OPTIMA 7 BIOGAS
10
2.3
Important general information (EN 50379) and VDI 4206
This analyzer is not designed to be used for continuous measurements.
Before using the analyzer verify the condition of the various parts of the analyzer, such as the probe, the ambient air conditions,
the condensate separator, pleated filter and the connectors for damage and/or blockages.
When starting up the analyzer it will take between 1 – 3 minutes to set to zero depending on the the sensors and ambient condi-
tions.
The minimum zeroing time of the analyzer to achieve correct measurement values can be expected by 1.5 minutes!
Caution:
Exposure to acids; aggressive gases such as sulphur; vapours such as thinners, gasoline, alcohol and paint, etc. can dam-
age, reduce the life of, or destroy the sensors.
The life span of the sensors depends on how they are used, maintained and treated. Typical average life expectations are: O
2
– 4-5
years; CO - 2 - 3 years and NO - 3 years.
The use of the analyzer for regulatory purposes is subject to special regulations (for example a periodical examination of the ana-
lyzer). Please obtain the appropriate regulations from your local responsible authority.
2.4
Important information about the users/operation manual
The users/operation manual is an important part of this delivery. It will explain how to use this analyzer properly and sets forth
safety and environmentally friendly procedues.
It is the responsibility of all users to read and familiarize themselves with this manual, paying particular attention to the safety
instructions.
The most important safety details are listed in chapter 2 (Safety Information).
Additional safety details in other chapters are clearly marked with an attention sign.
Содержание OPTIMA 7 BIOGAS
Страница 1: ...MRU Instruments Inc USER MANUAL OPTIMA7 BIOGAS 1...
Страница 39: ...MRU Instruments Inc USER MANUAL OPTIMA 7 BIOGAS 39 OK This user s decision must be confirmed see chap 16 2...
Страница 58: ...MRU Instruments Inc USER MANUAL OPTIMA 7 BIOGAS 58 16 7 Declaration of conformity...
Страница 59: ...MRU Instruments Inc USER MANUAL OPTIMA 7 BIOGAS 59...
Страница 60: ...MRU Instruments Inc USER MANUAL OPTIMA 7 BIOGAS 60 Technical data change w o notice Date of issue 20140731...