MRU Instruments, Inc.
USER MANUAL OPTIMA 7 BIOGAS
17
6.5
Condensate separator
Remove the condensate separator by pulling it
towards you (1) out of the groove of the OPTIMA 7
housing, then pull it downwards (2).
The condensate show glass is screwed into the
plug. The complete condensate separator can easi-
ly be taken apart for cleaning or for exchanging the
star filter. (Condensate separator details are in-
cluded in the service manual)
After cleaning a leak-tight test (see chap. 14.4) is
highly recommended!